Individual Test Description

1. RF IN Port VSWR Test

Description
This test checks the VSWR of the E5052A's RF IN port with a network analyzer. The
VSWR is checked at 10 MHz to 3 GHz.
Specification
@23 ± 5 °C
VSWR
Test equipment
Network Analyzer
Calibration Kit
N(m)-N(m) cable, 61 cm
Procedure
Step 1. Preset the Network Analyzer. Then set the Network Analyzer as follows.
Parameter
Measurement Mode
Cal Kit
Trigger Sweep
Sweep Type
Source Power
Data Format
Maker Display
Search Range
Chapter 2
Individual Test Description
< 1.6 @ 10 MHz to 30 MHz
< 1.2 @ 30 MHz to 2 GHz
< 1.3 @ 2 GHz to 3 GHz
Agilent E5070B/5071B
Agilent 85032F
Agilent N6314A(p/n 8120-8862)
Value
S11
85032F
Hold
Segment
0 dBm, ON
SWR
ON
ON
Performance Tests