7. DC Power Current Measurement Test

Description
This test checks the accuracy of the DC Power current measurement function. This test
uses a 141 Ω resistor to define the DC Power current.
Specification
@23 ± 5 °C
Accuracy
Test Equipment
Multimeter
Resistor
BNC cable, 61 cm, 2 ea.
Dual Banana-BNC(f) adapter
Procedure for Standard
Step 1. Connect the test equipment as shown in Figure 2-13
Figure 2-13
DC Power Current Measurement Test setup for Standard
Step 2. Press
of installing the firmware revision 1.00/1.10, press
E5052A settings
Measurement View :
FP-Current
Mode : Tester
Chapter 2
± (0.2 % of reading + 160 μA) @ 0 mA to 80 mA
[Preset]
-
Factory
to initialize the E5052A, then set the E5052A as follows. In case
Individual Test Description
Agilent 3458A
141 Ω, 1 W (Agilent p/n 0698-3615 (47 Ω), 3ea)
Agilent p/n 8120-1839
Agilent p/n 1251-2277
-
[Preset]
OK
Key strokes
[Meas/View]
-
Freq & Power
[Trigger]
-
Mode
-
Tester
Performance Tests
.
59