Power Measurement Accuracy Test - Agilent Technologies E5052A Service Manual

Signal source
Hide thumbs Also See for E5052A:
Table of Contents

Advertisement

Performance Tests
E5052A Performance Test Record
@ Frequency Band: 300 MHz - 7 GHz, Power Level: +20 dBm
Frequency
Frequency
[Hz]
Resolution [Hz]
300 M
64 k
300 M
1 k
300 M
10
7 G
64 k
7 G
1 k
7 G
10

Power Measurement Accuracy Test

@ -20 dBm
Frequency [Hz]
Test limit [dB]
1 G
2.0125 G
3.0125 G
4.0125 G
5.0125 G
6.0125 G
7 G
@ -10 dBm
Frequency [Hz]
Test limit [dB]
10 M
1 G
2.0125 G
3.0125 G
4.0125 G
5.0125 G
6.0125 G
68
Test Limit
[Hz]
± 64 k
± 1 k
± 10
± 64 k
± 1 k
± 10
Test result
± 1.0
± 1.0
± 1.0
± 1.0
± 1.0
± 1.0
± 1.0
Test result
± 1.0
± 0.5
± 0.5
± 1.0
± 1.0
± 1.0
± 1.0
Test result
[Hz]
[dB]
Measurement
uncertainty [dB]
± 0.13
± 0.14
± 0.15
± 0.16
± 0.16
± 0.17
± 0.18
[dB]
Measurement
uncertainty [dB]
± 0.14
± 0.13
± 0.13
± 0.13
± 0.16
± 0.16
± 0.17
Chapter 2

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents