Replacement Procedure

A4/A5 Local Synthesize Module Replacement

A4/A5 Local Synthesize Module Replacement

Tools Required

Removal Procedure

Refer to Figure 6-16 for this procedure.
Step 1. Remove the outer cover as described in "Outer Cover Removal" on page 176.
Step 2. Remove the front panel as described in "Front Panel Removal" on page 178.
Step 3. Disconnect the semi-rigid cables (item 1).
Step 4. Remove each two TORX T10 screws (item 2).
Step 5. Slide out the Local Synthesize module (item 3).
Figure 6-16
A4/A5 Local Synthesize Module Removal
Replacement Procedure
Step 1. Reverse the order of the removal procedure.
NOTE
Fasten the semi rigid cable using a torque wrench for replacement.
198
TORX screwdriver, T10, T15, and T20
Flat edge screwdriver
Open-end torque wrench, 5/16 inch (set to 10 in-lb) (for reconnecting SMA connector)