Performance Tests

E5052A Performance Test Calculation Sheet

E5052A Performance Test Calculation Sheet

Introduction

This section contains calculation sheets for each performance test that requires additional
calculations to determine the final test result.
Use the calculation sheet in this section as an aid for recording raw measurement data and
calculating the performance test results.
Calculation sheet entries are provided only for performance tests in which calculations are
required to obtain the test results.

SA Relative Level Accuracy Test

Sheet for Low Frequency (100 MHz)
Frequency [Hz]
reading [dBm]
SG1
SG2
PM1
100 M
92.5 M
93.5 M
94.5 M
95.5 M
96.5 M
97.5 M
98.5 M
99.5 M
100.5 M
101.5 M
102.5 M
103.5 M
104.5 M
105.5 M
106.5 M
107.5 M
64
Power meter
E5052A reading
[dBm]
PM2
SA1
Cal value
Test
of Step Att
Result
[dB]
[dB]
SA2
SATT
Result
Test result
equation
Result = (SA2 - SA1) -
(PM2 - PM1 - SATT)