Replacement Procedure

Front Panel Removal

Front Panel Removal

Tools required

Procedure

Refer to Figure 6-2 for this procedure.
Step 1. Remove the outer cover as described in "Outer Cover Removal" on page 176.
Step 2. Disconnect the two semi-rigid cables (item 1) from the front panel.
NOTE
When reconnecting the semi-rigid cables, make sure that the cables are connected to the
same connectors from which they previously came.
Step 3. Remove the top trim (item 2) using a flat edge driver.
Step 4. Remove the seven TORX T15 screws (item 3) fastening front panel.
Step 5. Gradually push the front panel assembly towards the outside.
Step 6. Release the cable (red and black) ( item 8) from the cable clamp under the FDD.
Step 7. Release the mylar cable and flat cables from the cable clamps (item 5).
Step 8. Disconnect the cables (item 4, 7) from the front panel.
NOTE
Keep the mylar cable (item 6) connected.
The mylar cable and its connector are very fragile. It's recommended that you should
replace the mylar cable when the cable is disconnected and reconnected frequently.
178
TORX screwdriver T10, T15 and T20
Flat edge screwdriver
Open-end torque wrench, 5/16 inch (set to 10 in-lb) (for reconnecting SMA connector)