Receiver Rf Flatness Adjustment; Receiver If Gain/Flatness Adjustment - Agilent Technologies E5052A Service Manual

Signal source
Hide thumbs Also See for E5052A:
Table of Contents

Advertisement

Adjustment
Required Adjustment after Replacing Assembly
Required equipment for the PM flatness adjustment
Description
Signal Generator
Power Splitter
Power Sensor
Power Meter
N(m)-N(m) Cable
N(m)-N(m) Adapter

Receiver RF Flatness Adjustment

The purpose of this procedure is to adjust the RF flatness of the A6 phase detector board.
Required equipment for the receiver RF flatness adjustment
Description
Signal Generator
Power Amplifier (35dB)
Power Amplifier (35dB)
50 ohm (BNC) Termination
Power Sensor
Power Meter
Power Splitter
N(m)-N(m) Cable
N(m)-N(m) Adapter

Receiver IF Gain/Flatness Adjustment

The purpose of this procedure is to adjust the IF gain / flatness of the A6 phase detector
board.
80
Recommended Model
Agilent E8257D option UNR,
UNT
Agilent 11667A
Agilent 8481A
Agilent E4419A/B
Agilent N6314A (p/n
8120-8862)
Agilent p/n 1250-1475
Recommended Model
Agilent E8257D option UNR,
UNT
R & K A3000-2H-R
Agilent 83020A
Agilent 11593A
Agilent 8481A
Agilent E4419A/B
Agilent 11667A
Agilent N6314A (p/n
8120-8862)
Agilent p/n 1250-1475
Chapter 3

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents