Adjustments For The New Hdd; Calibration Of The Touch Screen - Agilent Technologies E5052A Service Manual

Signal source
Hide thumbs Also See for E5052A:
Table of Contents

Advertisement

Replacement Procedure
A27 Mass storage Disk Drive Assembly Replacement

Adjustments for the New HDD

Step 1. Run the adjustment program.
Step 2. Input the GPIB address of the E5052A.
Step 3. Input the serial number of the E5052A.
Step 4. Perform the adjustment using "HDD" in Spot Adjustment of the program.

Calibration of the Touch Screen

When you have replaced the HDD on a E5052A, you have to calibrate the touch screen.
Follow the procedure described below to calibrate the touch screen.
Step 1. Press [System].
Step 2. Press
Step 3. Press
Step 4. Press
The touch screen calibration screen (Figure 6-4) appears.
Figure 6-4
Touch Panel Calibration Screen
Step 5. Touch the x mark on the upper left with your finger. The mark x appears also on the lower
left, upper right, and lower right. Touch the x marks in that order with your finger.
Touching the four locations described above with your finger automatically concludes the
touch screen calibration.
NOTE
With no operation on the touch screen calibration screen for a preset time, it automatically
closes and the previous measurement screen reappears.
182
Service Menu
.
.
Test Menu
Adjust Touch Screen
.
Chapter 6