M9210A Front Panel Features

Front Panel Connectors

Connector
INPUT 1
INPUT 2
TRIGGER IN
I/O A
CLK IN
I/O B
TRG OUT

Front Panel LED

STATUS
The maximum input to the input connectors (INPUT 1 and INPUT 2) must not exceed +27 dBm into
50 Ω, or 300 V into 1 MΩ.
Agilent M9210A PXI-H 10-bit Digitizing Scope Startup Guide
Description
This BNC female connector is the analog signal input 1. It features a switchable
50 Ω/1 MΩ input stage with programmable voltage ranges from 50 mV to 50 V full
scale (in a 1, 2, 5 sequence) with variable voltage offset and a bandwidth of
1.4 GHz/300 MHz (respectively 50 Ω/1 MΩ).
This BNC female connector is the analog signal input 2. It features a switchable
50 Ω/1 MΩ input stage with programmable voltage ranges from 50 mV to 50 V full
scale (in a 1, 2, 5 sequence) with variable voltage offset and a bandwidth of
1.4 GHz/300 MHz (respectively 50 Ω/1 MΩ).
This BNC female connector is a 50 Ω terminated input for an external trigger sig-
nal. The trigger frequency range is DC to 1 GHz, full scale (FS) is 0.5/1/2/5 V and
threshold adjust range ± FS/2.
This MMCX female connector provides access to an I/O digital control line for mon-
itoring or modifying the Digitizing Scope status and configuration or extracting a
10 MHz clock reference signal.
This MMCX female connector is used for the external reference or the external
clock signal input . The external reference clock frequency is 10 MHz. The exter-
nal clock signal frequency range is 200 MHz to 2 GHz. In either case the ampli-
tude should be 500 mV pk-pk or higher. Maximum DC voltage should not exceed ±
5 V.
This MMCX female connector provides access to an I/O digital control line for mon-
itoring or modifying the Digitizing Scope status and configuration or extracting a
10 MHz clock reference signal.
This MMCX female connector features a trigger signal output.
LED
(off)
Power is not present or the module has a fault.
Indicates that the trigger is armed and waiting for a valid trigger to
(green)
occur.
(red)
Indicates that a trigger has occurred, and the acquisition is complete.
The power-on state immediately after power is applied. Identifies the
(yellow)
module associated with the current channel in the Soft Front Panel
when the 'Identify' function is used.
M9210A Front Panel Features
Description
13