Printout Configuration

Purpose
This menu item enables you to configure
individual printout formats. With the
formulation, totalizing and statistics
applications, you can also define the
values to be included on the total
printout when the MR key is pressed.
Under "Setup: Printout: Application-
defined output: Configured printout", you
can configure individual, component or
total data records that contain the items
available for printouts in each application.
Configure these printouts after you have
configured the applications, because some
entries in the data record depend on the
particular application.
32
Features
– Maximum items in a data record: 60
– Separate configuration of printout
formats for individual weights,
components, total, backweighing
and statistics
– Individual printout generation:
press the p key
Automatic printout of application data:
results from animal weighing or density
application (Setup menu: Application 1:
Density: Printout: All data) OK values
from checkweighing application, time-
controlled printouts, 2nd tare memory
– Component printout:
For results from totalizing, formulation
or statistics applications, press
M-
(Setup: Application 3: ..., Printout
of individual components: On)
– Total printout:
For totalizing, formulation or statistics
MR
applications, press
– Backweighing printouts or records:
automatically generated after back-
weighing or manually by pressing the
p key when the result is displayed
at the end of backweighing
– Statistics printout or output:
To generate, press the p key when
the statistics are displayed
Printouts for Differential Weighing:
These printouts can be generated as
standard or configured (user-defined)
reports.
You can configure the following
printouts:
– Individual printout
– Backweighing printout
– Statistics printouts
Printouts are generated in one
of two ways:
– at the request of the user by pressing
the p key (print on request)
– automatically, if configured in the
Setup menu [Application parameters:
Application 1: Differential weighing:
Generate printout: Auto]
You can turn off automatic printout
generation in the Setup menu
[Application parameters: Application 1:
Differential weighing: Generate
printout: None]
Data records are deleted after you have
switched to a different application or
activated or de-activated an extra
function in the application parameters
of the Setup menu
– A new pick list for a data record is
created based on the currently active
application programs and extra
functions
– Printout items can be deleted
individually
M+
or
– No printout is generated when the
following setting is configured:
Setup: Printout: Line format:
For raw data (16 characters)
– Print item "Form feed" for footer:
Advance to beginning of next label
in the "YDP01IS-Label" and
"YDP02IS-Label" interface mode
Extra Functions
§ Exit printout configuration:
press
soft key
oo
> Restart application
Printing "Select" and "List" Settings
LIST
: print the currently selected
Select
list
can still be selected
§ When the select bar is on
Select
or
> Printout (Example)
BACKW. PRINT.LIST
====================
Sample date
Net initial wt.
Backweighed res
Loss in %
====================
etc.