Setting the Scale Functions (

Purpose
This menu item enables you to
configure the scale functions, i.e.,
to meet individual requirements by
selecting predefined parameters in the
Setup menu. You can block access
to the menu by assigning a password.
Features
The scale functions are combined in
the following groups (1st menu level):
– Calibration/adjustment
– Adapt filter
– Application filter
– Stability range
– Taring
– Auto zero
– Weight unit 1
– Zero range
– Zero range at power on
– Tare/zero at power on
– Factory settings: only wgh. param.
(only the scale functions)
Factory Settings
Parameters: The factory settings are
identified by the symbol "ο" in the list
starting on the next page.
BAL.FUNC.
Preparation
Show available scale functions:
§ Select Setup menu: press the H key
> SETUP is displayed
§ Select "Balance/scale functions":
press the
soft key
O
If you already assigned a password:
> The password prompt is displayed
$ If access is blocked by a password:
enter the password using the numeric/
alphabetic keys.
$ If the last character of the password is a
letter: conclude input by pressing a
§ Confirm your password and have
the scale functions displayed:
Press the
soft key
l
> Scale functions are displayed:
$ To select the next group:
press the
soft key (down arrow)
q
$ To select the previous item of a group:
press the
soft key (up arrow)
Q
$ To select the next sub-item within
a group: press the
soft key
O
(right arrow)
$ To select the previous group:
press the
soft key (left arrow)
o
$ To confirm: press the
soft key
l
)
Extra Functions
§ Exit the Setup menu:
press the
oo
> Restart your application
§ Print parameter settings:
– When the balance/scale functions are
displayed, press p
> Printout (example)
Texts with more than 20 characters
are cut off
SETUP
BAL.FUNC.
---------------------
Calibration/adjustm
CAL/iso TST key fun
Internal cal./adju
Cal/adjustm seq
Cal. with adjustm au
.
isoCAL-function
Start autom. adjus
Print GLP/GMP adju
Automatic if GLP is
selected
Parameter for exte
Wt. ID (W ID):
Cal./adjust.-wt:
Adapt filter
Normal vibration
Application filter
Stability range
2 digits
Taring
After stability
Auto zero
Weight unit 1
etc.
soft key
Off
isoCAL
5000.00 g
Filling mode
Off
Grams /g
17