Display/Problem
Err 31
Print fct. blocked
Ref.wt. too light
Cannot update
Not a number
Too low
xxxxx
Too high
xxxxx
Too many char.
Wrong line format
Limits unequal
for unit
Equation too long
Err 10
x
x = 1 :
x = 2 :
x = 3 :
x = 4 :
"Checkerboard" pattern displayed
continuously
Err 320
Err 340
Err 341
No WP
blocked
The special code
J
remains displayed
The weight readout
changes constantly
The weight readout is
obviously wrong
If any other errors occur, please contact your local Sartorius Service Center.
Cause
Interface handshake interrupted
(XOFF, CTS)
Error in storing reference weight
(with the counting or weighing-
in-percent application)
Reference updating not possible
(with the counting application)
Input wrong (with any application pro-
gram), e.g., alphabetic input not allowed
Input text too long
Configured printout, printout memory
and 16-character format selected
Unit entered for tolerance limits in check-
weighing different from the appl. used
Equation exceeds 28 characters
in formulation
Key is stuck
Key pressed when switching on the scale:
(F1, F2, F5, F6), c
D,
(F3), 0, 3, 4, 9
2, 5, 6, ., p, t-right
1, 7, 8,
(F4), a, t-left
H key was pressed when turning on
the scale, or is stuck
Operating program memory defective
Operating parameter (EEPROM)
is wrong
RAM lost data
Factory settings deleted
Battery needs to be recharged
Weighing platform is defective
Function blocked
None of the keys has been pressed
since the scale was turned on
Unstable ambient conditions
Too much vibration, or the scale
is exposed to a draft
A foreign object is caught between
the pan and the scale housing
The scale has not been
calibrated/adjusted
The scale was not tared before weighing
The scale is not level
The dust cover is caught under the
weighing pan
Solution
Transmit XON, then CTS
Weight too light or there is no sample
on the scale
See "Counting" in "Operation" for reference
updating criteria
Follow the instructions for the application
programs
Allowable text lengths, incl. decimal point:
– S ID, NUM, L ID,
ID: max. 20 characters
– W ID: max. 14 characters
Select the 22-character format
Adjust tolerance limits
Limit equation to 28 characters
Release key or
Contact your local Sartorius
Service Center
Contact your local Sartorius
Service Center
Turn the scale off, then back on
again. If this error remains displayed,
please contact your local Sartorius
Service Center
Leave the scale power on for at
least 10 hours
Contact your local Sartorius Service Center
None
Press a key
Set up the scale in another area
Change Setup configurations to adapt
the scale to the ambient conditions
Remove the foreign object
Calibrate/adjust the scale
Tare before weighing
Level the scale
See "Replacing the Dust Cover" in
the chapter "Care and Maintenance"