Printout Configuration

Purpose
This menu item enables you to
configure individual printout formats.
With the formulation, totalizing and
statistics application, you can also
define the values to be included on
the total printout when the MR key
is pressed.
Under "Setup > Printout > Application-
defined output > Configured printout",
you can configure individual, compo-
nent or total data records that contain
the items in each application that are
available for printouts. Configure these
printouts after you have configured the
applications, because some entries in
the data record depend on the particu-
lar application.
36
Features
– Maximum items in a data record: 60
– Separate configuration of printout
formats for individual weights,
components, total, backweighing and
statistics
– Individual printout generation:
p
press the
/r key
Automatic printout of application data:
e.g., results from animal weighing or
density application (Setup menu:
Application 1: Density: Printout: All
data) OK values from checkweighing
application, time-controlled printouts,
2nd tare memory
– Component printout:
For results from totalizing, formulation
or statistics applications, press
M+
M-
or
(Setup: Application 3: ...,
Printout of individual components: On)
– Total printout: For totalizing,
formulation or statistics applications,
MR
press
– Backweighing printouts or records:
automatically generated after back-
weighing or manually by pressing the
p
/r key when the result is dis-
played
at the end of backweighing
– Statistics printout or output:
p
To generate, press the
when the statistics are displayed
Printouts for Differential Weighing:
These printouts can be generated as
standard or configured (user-defined)
reports.
You can configure the following
printouts:
– Individual printout
– Backweighing printout
– Statistics printouts
Printouts are generated in one of
two ways:
– at the request of the user by pressing
p
the
/r key (print on request)
– automatically, if configured
in the Setup menu [Application
parameters: Application 1: Differential
weighing: Generate printout: Auto]
You can turn off automatic printout
generation in the Setup menu
[Application parameters:
Application 1: Differential weighing:
Generate printout: None]
– Data records are deleted after you haved
switched to a different application or
activated or de-activated an extra
function in the application parameters
of the Setup menu
– A new pick list for a data record is
created based on the currently active
application programs and extra
functions
– Printout items can be deleted
individually
– No printout is generated when the
following setting is configured:
Setup: Printout: Line format:
For raw data (16 characters)
– Print item "Form feed" for footer:
Advance to beginning of next label in
the "YDP01IS-Label" and "YDP02IS-
Label" [printer] interface mode
Extra Functions
§ Exit printout configuration:
press
oo
> Restart application
/r key
Printing "Select" and "List" Settings
LIST
: print the currently selected list
SELECT
be selected
§ When the select bar is on
LIST
or
press the
> Printout (Example)
BACKW. PRINT.LIST
====================
Sample date
Net initial wt.
Backweighed res
Loss in %
====================
etc.
soft key
: printout items that can still
SELECT
:
p
/r key