Operation; Basic Weighing Function - Sartorius FC06BBE-S Operating Instructions Manual

Factory series fc and fcg models electronic precision scales
Hide thumbs Also See for FC06BBE-S:
Table of Contents

Advertisement

Operation

Basic Weighing Function

Purpose
The basic weighing function is always
accessible and can be used alone or
in combination with an application
program (Toggle between Weight Units,
Counting, Weighing in Percent, etc.).
Features
– Taring the scale
– Assigning IDs to weights
– Printing weights
– Printing ID codes for weights
Factory Settings
After stability
Tare:
Manual/auto print mode:
Manual with stability
Line format:
For other apps/GLP
(22 characters)
Soft Key Functions
isoTST
Initiate calibration/
adjustment routine
isoCAL
Press to start isoCAL
routine
S ID
Store ID entered
38
Under-Scale Weighing
A port for a under-scale weighing
hanger is located on the bottom of
the scale.
FC... BBE, FC... CCE:
§ Open cover plate (1) on the bottom
of the scale
1
§ Attach the sample (e.g., using a
suspension wire) to the hook (2).
2
FCG... EDE:
§ Use a suitable screwdriver to remove
the cover plate from the bottom of
the scale.
§ Fasten the hook to the cross-bar
(see "Accessories").
$ If necessary, install a shield for protec-
tion against drafts
!
Important Note Concerning Verified
Scales Approved for Use as Legal
Measuring Instruments in the EU*:
The under-scale weighing port may not
be opened or used when an approved
scale is being operated as a legal
measuring instrument.
Preparation
§ Turn on the scale: Press e
> The Sartorius logo is displayed
$ To tare the scale, if desired: Press t
> The
symbol is displayed when a
S
verified scale is tared or zeroed
(±0.25 digits).
For Service Personnel:
Important Note Concerning Verified
Scales Approved for Use as Legal
Measuring Instruments in the EU*:
The type-approval certificate for verifi-
cation applies only to non-automatic
weighing instruments; for automatic
operation with or without auxiliary
measuring devices, you must comply
with the regulations of your country
applicable to the place of installation
of your scale.
§ You must calibrate the scale at the
place of installation before using it as
a legal measuring instrument (see the
section entitled "Calibration and
Adjustment" in this chapter)
$ The temperature range indicated on
the verification ID label must not be
exceeded during operation
* including the Signatories of the
Agreement on the European
Economic Area