Counting
A
Preparation
To calculate a piece count, the average weight of one piece must be known.
This average piece weight can be entered into the Counting program in one of three
ways:
– Enter the average piece weight using the numeric keys and store it;
– The last reference sample quantity entered is loaded and displayed when you turn on
the scale. Place the same number of parts on the balance and initialize the Counting
program;
– When the automatic initialization parameter (see previous page), is on (Setup: Printout:
Application-defined: Autoprint upon initialization: All values; see page 34), the scale
goes into the "counting" mode when you turn it on and loads the last average piece
weight and corresponding reference sample quantity that were entered or calculated.
Reference Sample Updating
You can have the average piece weight updated during counting (with the piece count
displayed) if "AWP update" is set to "manual" or "automatic" in the Setup menu.
Manual updating can only be performed when the
Reference sample updating must be completed before using an application program
from Application 3.
Opt.
The
soft key is displayed when:
– the scale has stabilized (stability symbol displayed)
– the current piece count is not more or less than double the original piece count
– the current piece count is more than 100
– the internally calculated piece count (e.g., 17.24) differs from the nearest whole number
(in this case: 17) by less than ±0.3
Reference updating can be repeated several times with an approximately doubled piece
count.
$ To perform reference updating: press the
$ Turn on the scale: press e
> Sartorius logo is displayed
Select the Counting application in the Setup menu: press H
Application parameters
Select the
key once
Application 1 (basic settings)
Select
Counting
Select
: repeatedly press the
Counting
Confirm
: press the
Counting
Average piece
weight calculation
Average piece
weight updating
ο = factory setting
see also the "Application Parameters (Overview)" in the chapter entitled "Configuration"
Save settings and exit the Setup menu: press the
56
Update
soft key is displayed.
Opt.
soft key
: press the
key 2+, then the
q
: press the
or
soft key
Q
q
soft key
O
ο Display accuracy
+1 decimal place
+2 decimal places
Off
Manual
ο Automatic
soft key
oo
Additional Functions
In addition to the functions for:
– alphanumeric input,
– taring (not during alphanumeric input),
– printing,
you can also access the following
functions from this application:
Calibration/Adjustment
Press the
> See "Calibration/Adjustment" for
further instructions
Toggling to the Next Application
Press d
> See the section on the corresponding
application program for further
instructions
Setup (Setting Parameters)
Press H
> See "Configuration" for further
instructions