Sartorius PMA.Quality PMA7501-Y Installation Instructions Manual

Sartorius PMA.Quality PMA7501-Y Installation Instructions Manual

Advertisement

Quick Links

Installation Instructions
Sartorius PMA.Quality
Model PMA7501-Y | -Y00W | -Y00U
Electronic Paint-mixing Scale for Use in Zone 2
Potentially Explosive Atmospheres
98648-014-52

Advertisement

Table of Contents
loading

Summary of Contents for Sartorius PMA.Quality PMA7501-Y

  • Page 1 Installation Instructions Sartorius PMA.Quality Model PMA7501-Y | -Y00W | -Y00U Electronic Paint-mixing Scale for Use in Zone 2 Potentially Explosive Atmospheres 98648-014-52...
  • Page 2: General View Of The Equipment

    General View of the Equipment 1 Display and control unit 10 Display 2 Q key (On/Standby) 11 Weighing pan 3 R key: Upwards 12 Interfaces (D-Sub plug, 9-contact) 4 S key: Downwards 13 Connection to AC power 5 U key: Zero/Tare 14 Grounding terminal 6 I key (TOGGLE) 15 Column...
  • Page 3: Table Of Contents

    Make sure the equipment is accessible from a paint manufacturer) installed on at all times. a connected PC. If you wish to create your own application software, Sartorius General Provisions for Installing can supply the required drivers for DOS or The PMA7501-Y models with power ®...
  • Page 4 Sartorius before connecting the cable to – Unless performed by an authorized Sartorius equipment, and disconnect any Sartorius technician, any installation work wires that are assigned differently.
  • Page 5 IP Protection – Use original Sartorius spare parts, only! – Handle the equipment and any – Never use a hammer to close the lid of a accessories in accordance with the IP paint can while it is still on the weigh- rating (EN 60529).
  • Page 6: Getting Started

    Getting Started § Remove the scale from its packaging. § After unpacking the scale, check it immediately for any visible damage as a result of rough handling dur- ing shipment. Equipment Supplied – Scale – Weighing pan – Power supply YPS04-Y.. Setting Up the Scale Choose a suitable place to set up the scale.
  • Page 7 Sartorius Customer Service for information on the legal regulations applicable in your country. Use only genuine Sartorius power supplies. The use of power supplies from other manufacturers, even if these units have a registered approval rating from a national testing laboratory, requires the approval of a certified technician.
  • Page 8: Operating The Equipment

    Operating the Equipment Turn on the scale using the After the scale has been If a different readout is Q key (2). turned on, it will automati- displayed, zero or tare the cally run a self-test. scale using the U key (5). At the end of this test, 0.0 g is displayed.
  • Page 9 Weighing with Two Decimal Places Note: To weigh using two decimal places, you must first adapt the settings (refer to the chapter entitled “Menu Settings”.) Press the I key (6). The display shows “0.00 g.” Place an empty paint can on the weighing pan (11). Press the U key (5).
  • Page 10: Applications

    Applications Formulation Mode (Calculation by a Factor) This mode enables you to weigh in amounts that are smaller or larger than that of your basic formula for a specific paint color (e.g., 250 ml of a 1-l formula). You can select various factors (amounts) by pressing the K formulation key (7): 0.25 0.5 0.75 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0.
  • Page 11 3. “.25” is displayed next to the weight 4. Slowly pour in the first component, “250 g” of green paint, until the display shows “250 g.” 5. Pour in the second component, “250 g” of red paint, until the display shows “500 g.” 6.
  • Page 12 Example (kumulativ): Place an empty paint can on Press the U key (5) Pour in the first component. the weighing pan (11). 0.0 g + 50.0 g + 118.0 g Press the L key [MEM] (9). Pour in the 2nd component. Press the L key [MEM] (9).
  • Page 13: Calibration/Adjustment

    Calibration/Adjustment You can calibrate/adjust the scale by pressing the U key (5). Calibration weight: 5,000 g; accuracy: + 0.075 g. After connection to AC power and before each calibration/adjustment, allow the scale to warm up for approx. 30 min. Hold down the U key (5) for 2 sec. When 5000 is displayed, release the key.
  • Page 14: Menu Settings

    § Press the L key [ENTER]. “o” will appear. The new code is stored. § Press the c key (Clear) several times to exit the menu. Note: To obtain a detailed list of the menu codes, please ask your nearest Sartorius office.
  • Page 15 Important Menu Settings § Hold down the L key [ENTER] for approx. 2 sec. “SETUP” will appear on the display (Level 1). Level 1 SETUP Language Settings Level 1 Level 2 Level 3 Level 4 § R key: select “LANGUAGE” LANGUAGE §...
  • Page 16 Toggling Decimal Places (Standard = 1 decimal place PolyRange = 2 decimal places) Toggling Units (Grams or PT./PD.) These settings are active when the scale is switched on. Level 1 Level 2 Level 3 Level 4 § Press the L key [ENTER] SETUP §...
  • Page 17 Note: Keep the numeric code in a safe place. The scale can only be accessed by entering the correct code. Level 1 Level 2 Level 3 Level 4 § RS keys: select “INPUT” INPUT § Press the L key [ENTER] PASSWORD §...
  • Page 18 § Press the L key [ENTER]; “o” will appear: etc. the new code is set § Press the c key (Clear) several times o CODES to exit the menu. Note: To obtain a detailed list of the menu codes, please ask your nearest Sartorius office.
  • Page 19: Troubleshooting

    Troubleshooting Problem Cause Solution No segments appear – No AC power available – Check the AC power supply on the weight display Weight display shows – The weighing pan is – Position the weighing pan “Low” is not in place Weight display shows –...
  • Page 20: Care And Maintenance

    Observe the Warning and Safety been packaged using environmentally- Information friendly materials. In this case, notify your nearest Sartorius You should retain these materials in case Service Center or the International Tech- you need to package your scale for nical Support Unit based in Goettingen, storage or return shipment.
  • Page 21: Recycling

    EU: 1) Go to http://www.sartorius.com. 2) Select the summary under “Service.” 3) Then select “Information on Disposal.” 4) Addresses for local Sartorius disposal contacts can be found in the PDF files given on this webpage.
  • Page 22: Specifications And Interface Ports

    Specifications and Interface Ports PMA 7501-Y | PMA 7501-Y 00W Weighing range 999.95/7500 Readability 0.05/0.1 Tare range (subtractive) –999.95/–7500 Max. linearity < ± 0.2 Stability range digit 0.25 to 4 Moisture-proof rating Non-condensing Allowable ambient operating temperature range °C 0 to +40 Weighing pan d mm Scale housing (W + D + H)
  • Page 23 PMA 7501-Y00U Weighing range 999.95/7500 Readability 0.05/0.1 Tare range (subtractive) –999.95/–7500 Max. linearity < ±0.2 Stability range digit 0.25 to 4 Moisture-proof rating Non-condensing Allowable ambient operating temperature range °C 0 to +40 Weighing pan d mm Scale housing (W + D + H) 233 + 329 + 391 Net weight, approx.
  • Page 24: Accessories

    Accessories Order no.: YDC01PMA In-use dust cover YCC01-0027M2 RS-232 Data cable (SBI) (2 m) YCC01-0028M2 RS-232 Data cable (BPI) (2 m) YCC01-0028M20 RS-232 Data cable (BPI) (20 m) YCO12 USB/RS-232 Data cable (SBI) (1.80 m) YCO13 USB/RS-232 data cable (BPI) (1.80 m) YCC01-0040M3 USB data cable...
  • Page 25: Usb Port (Pc)

    USB Port (PC) Purpose Any PMA7501-000U can be connected to a PC equipped with a USB port. A virtual serial interface (virtual COM port) is set up as a device type at the USB port. This virtual serial interface is identified und operated by the application program. The protocols xBPI and SBI can be transmitted via the USB port.
  • Page 26 Installation Guides for Windows XP ® , Windows Vista ® and Windows 7 ® Changing the Port Number If you use the USB interface with a program that limits the number of COM port designations (e. g., only COM1, 2, 3, 4), you may have to assign one of these port numbers to the new virtual port.
  • Page 38 Sartorius Lab Instruments GmbH & Co. KG Otto-Brenner-Strasse 20 37079 Goettingen, Germany Phone: +49.551.308.0 Fax: +49.551.308.3289 www.sartorius.com The information and figures contained in these instructions correspond to the version date specified below. Sartorius reserves the right to make changes to the technology, features, specifications and design of the equipment without notice.

This manual is also suitable for:

Pma.quality pma7501-y00wPma.quality pma7501-y00u