Step
9. Store piece count
10. Unload the scale
11. Place another load of parts
on the scale (e.g., 29 pcs)
12. Add piece count to stored total
13. Repeat steps 11 and 12 as required
14. Display final evaluation ("Info" window)
(here: 5 weighing operations;
total quantity: 165)
The o indicates which value is
displayed in the text line; you can
change this selection
15. Print final evaluation
90
Key (or instruction)
M+
soft key
Remove parts from the scale
Place parts on the scale
M+
soft key
MR
soft key
p
Display/Output
6
isoTST
--------------------
16.01.1997
n
Qnt
+
6
isoTST
6
isoTST
n
+
Qnt
+
--------------------
n
Mittel+
s
+
srel
+
Summe +
Min
+
Max
+
Diff
+
16.01.1997
--------------------
11:06
1
35 pcs
2
29 pcs
5
33.0 pcs
3.2 pcs
9.70 %
165 pcs
29 pcs
37 pcs
8 pcs
11:16