Setting The Balance Functions - Sartorius ME215 Operating Instructions Manual

Me series se series electronic analytical balances and semi-micro-, micro- and ultra-microbalances
Table of Contents

Advertisement

Setting the Balance Functions (

Purpose
This menu item enables you to
configure the balance functions, i.e.,
to meet individual requirements by
selecting predefined parameters in the
Setup menu. You can block access to
the menu by assigning a password.
Features
The balance functions are combined in
the following groups
(1st menu level):
– Calibration | adjustment
– Adapt filter
– Application filter
– Stability range
– Taring
– Auto zero
– Weight unit 1
– Display accuracy 1
– Tare/zero with power on
– Factory settings: only wgh. param. (only
the balance functions)
For legal metrology, the selection
of individual parameters is limited.
Factory Settings
Parameters: The factory settings are
identified by the symbol "o" in the list
starting on page 27.
26
BAL.FUNC.
Preparation
Show available balance functions:
§ Select Setup menu: press the o key
> SETUP is displayed
§ Select "Balance functions":
press the
soft key
O
If you already assigned a password:
> The password prompt is displayed
$ If access is blocked by a password:
enter the password using the numeric |
alphabetic keys
$ If the last character of the password is
a letter: conclude input by pressing a
§ Confirm your password and have
the balance functions displayed:
Press the
l
soft key
> Balance functions are displayed:
$ To select the next group:
press the
soft key (down arrow)
q
$ To select the previous item of
a group: press the
soft key
Q
(up arrow)
$ To select the next sub-item within
a group:
press the
soft key (right arrow)
O
$ To select the previous group:
press the
soft key (left arrow)
o
$ To confirm: press the
soft key
l
)
Extra Functions
§ Exit the Setup menu:
press the
soft key
oo
> Restart your application
§ Print parameter settings:
– When the balance functions are
displayed, press r
> Printout (example)
Texts with more than
20 characters are cut off
SETUP
BAL. FUNC.
--------------------
Calibration/adjustm
CAL/isoTST key fun
Selection mode
Cal/adjustment seq
Calibrate, then auto
isoCAL function
On without resetting
Start automatic ad
Print GLP/GMP adju
Automatic if GLP is
Parameter for exte
Wt.
ID (W ID):
Cal./adj. wt:
200.00000 g
Adapt filter
Normal vibration
Application filter
Final readout
Stability range
Auto zero
Weight unit 1
Display accuracy 1
--------------------
adjust
app.
isoCAL
selected
2 digits
On
Grams /g
All digits