Ai Measurement - Omron HEM-9000AI Instruction Manual

Non-invasive blood pressure monitor with augmentation index (ai)
Hide thumbs Also See for HEM-9000AI:
Table of Contents

Advertisement

14.AI Measurement

14.AI Measurement
14.1. AI Measurement
This mode measures only AI.
1.
Attach the AI pulse wave sensor unit to the left wrist. (Refer to Section 9.2.)
2.
Check that the patient's name and ID appear in the
upper-left part of the screen. [Fig. 1]
If nothing is displayed, follow the instructions in Section
17. to register or retrieve the ID.
If the ID information is not registered, only the time and
results will be saved. Check that the measurement
method in the lower-left part of the screen is set to AI
measurement.
* Do not press the START button without first attaching the AI
pulse wave sensor unit.
* Do not slide the AI pulse wave sensor unit while AI
measurement is in progress.
3.
Press the START button. [Fig. 2]
The AI pulse wave sensor unit will begin applying
pressure.
To stop the device once measurement has begun, press
the STOP button. Rapid deflation will occur.
If measurement takes more than five minutes, the device
will stop automatically.
Pressing the DISPLAY button during measurement will
stop the measurement and turn off the screen.
4.
Check that the tonogram in the upper-centre part of
the screen is rising and forming a mound shape.
[Fig. 3]
(Refer to Section 12.1.)
(A continuous pulse wave is displayed on the graph in the
lower-left part of the screen.)
40
Continuous Pulse Wave Graph
[Fig. 1]
[Fig. 2]
Tonogram
[Fig. 3]