3.
Turn the patient's right palm up. [Fig. 5]
Always wrap the cuff around the upper part of the
patient's right arm directly on the skin. Wrapping the
cuff around bulky clothing or a rolled-up sleeve can
cause the readings to be inaccurate.
4.
Align the brachial artery with the artery position mark
"
". [Fig. 6]
5.
Wrap firmly with both hands, snug, and make sure to fasten with the hook and loop
fastener. For this step, place the cuff 1 ~ 2 cm (a little less than one inch) above the inside
of the elbow joint. [Figs. 7-10]
* If after wrapping, INDEX, printed on the cuff, is misaligned with RANGE, select a different, more
appropriate size cuff according to the circumference of the arm, and wrap again.
* If not wrapped with appropriate cuff, the arm might become numb, or the HEM-9000AI might not be able to
measure blood pressure correctly.
* Wrapping the cuff at an angle will have no impact on the measurements, so allow the cuff to follow the
shape of the arm.
[Fig. 7]
[Fig. 9]
6.
Align the cuff to be at the same height as the heart.
* Maintain the height of the cuff at the same height as the
heart in order to measure accurately.
[Fig. 11]
[Fig. 5]
[Fig. 6]
[Fig. 8]
1~2 cm
[Fig. 11]
9.Measurement Preparation
[Fig. 10]
25