Omron HEM-9000AI Instruction Manual page 88

Non-invasive blood pressure monitor with augmentation index (ai)
Hide thumbs Also See for HEM-9000AI:
Table of Contents

Advertisement

21.Error List
Error Codes for AI Pulse Wave Measurement
Error
Error Name
No.
AI Pulse Wave
E50
Sensor Unit
Malfunction.
AI Pulse Wave
E51
Sensor Unit
Disconnected.
AI Pulse Wave
E52
Sensor
Malfunction.
AI Pulse Wave
E53
Sensor Error.
AI Pulse Wave
E54
Sensor Error.
AI Pulse Wave
E55
Sensor Error.
AI Pulse Wave
E56
Sensor Unit
Malfunction.
Air Leakage
E60
Error.
E61
Deflation Error. The cuff was not
Rapid Exhaust
E62
Error.
Inflation Pump
E63
Error.
Intake Pump
E64
Error.
AI Pulse Wave
Measurement
E65
Unit Sensor
Error.
88
Error Message on
Display
Abnormality is
detected in the
output of the AI
pulse wave sensor
unit.
The signal from the
An error has occurred in
AI pulse wave
the AI pulse wave
sensor unit cannot
sensor unit.
be detected.
Abnormality is
detected in the
output of the AI
pulse wave sensor
unit.
The pulse wave
An error has occurred in
cannot be accurately
the AI pulse wave
detected.
sensor unit.
An error has occurred in
Abnormality is
the AI pulse wave
detected in the
sensor unit.
output of the AI
pulse wave sensor
unit.
There is a circuit failure
in the AI pulse wave
sensor unit.
There is air leaking
in the AI pulse wave
measurement unit.
An error has occurred in
deflated properly.
the AI pulse wave
Pulse wave was not
measurement unit
measured.
pressure control
system.
The cuff was not
inflated properly.
Pulse wave was not
measured.
An error has occurred in
There is air leaking
the AI pulse wave
in the AI pulse wave
measurement unit
measurement unit.
pressure control
system.
An error has occurred in
AI pulse wave
the AI pulse wave
measurement unit
measurement unit
malfunction.
pressure control
system.
Error Content
• Turn the power off, turn it back
on and check the pulse wave
sensor unit. If this does not
solve the problem, contact
OMRON.
• There is a possibility the AI
pulse wave sensor is not on the
artery or that the wrist belt is
loose. Reattach the AI pulse
wave sensor unit properly.
Refer to Section 9.2. Retake the
measurement.
• Turn the power off and check
the pulse wave measurement
unit connection. If the error
cannot be solved by the
suggested action, contact
OMRON.
• Turn off the main unit.
• Turn off the power and check
the AI pulse wave
measurement unit air tube
connection.
• Check the AI pulse wave
measurement unit air tube
connection.
• Turn off the power and check
the AI pulse wave
measurement unit air tube
connection.
Action