5.
As the AI pulse wave sensor pressure becomes
stable, the pulse wave will be displayed in the upper
left corner of the screen, and the actual
measurement will begin. [Fig. 5]
Instruct the patient not to move during
measurement. A measurement error will occur as a
result of body movement or vibration.
If the AI value goes outside of the 30%-150% range,
the AI measurement results will not be displayed for
that beat. Check to make sure that there is no
abnormality in the pulse waveform. This will also
cause a break in the graph of the AI values plotted
for each pulse wave.
If the pulse is irregular or too faint, the device may
not detect the pulse wave, in which case there will
be no pulse waveform and nothing will be displayed
on the graph.
A green bar in the lower-left part of the screen shows
the measurement progress toward completion.
6.
Once measurement is complete the results will
appear, along with a light blue "complete" message
at the lower left corner of the screen. Next, the green
message [SAVING] will be displayed at the lower left
corner of the screen. Measurement data will be
stored in the memory card. [Fig. 6]
Do not insert or remove the memory card during the
save operation. This will result in device failure.