INSTRUCTION MANUAL
IntelliSense
Blood Pressure Monitor
®
HEM-907XL
Model
Thank you for purchasing the OMRON
IntelliSense
®
Blood Pressure Monitor.
Please thoroughly read this Instruction Manual
before using this monitor to ensure safe and
accurate use.
Please keep this manual near the monitor all
the time for future reference.

Table of Contents

Be Sure to Read This Section
Exemptions.........................................2
®
Notes on Safety..................................3
Features of the Product .....................7
Components of the Product ...............8
Features and Functions .....................9
Names of the Parts ........................9
while the unit is in use ..................11
How to Apply the Cuff ......................13
of Battery Pack .............................17
SINGLE Mode ..............................19
AVG Mode ....................................21
MANU Mode .................................23
List of Error Codes ...........................27
Troubleshooting ................................28
Specifications
Caution.............................................29
Specifications ...................................30
16
...........

Advertisement

Table of Contents
loading

Summary of Contents for Omron INTELLISENSE HEM-907XL

  • Page 1: Table Of Contents

    INSTRUCTION MANUAL IntelliSense Blood Pressure Monitor ® HEM-907XL Model Thank you for purchasing the OMRON IntelliSense ® Blood Pressure Monitor. Please thoroughly read this Instruction Manual before using this monitor to ensure safe and accurate use. Please keep this manual near the monitor all the time for future reference.
  • Page 2: Exemptions

    EXEMPTIONS OMRON does not accept liability and warranty becomes void under the following circumstances: 1. When persons, not authorized by OMRON, perform repairs or modifications of this product. 2. When use and/or operation of this device is adversely effected by a device not manufactured by OMRON.
  • Page 3: Notes On Safety

    NOTES ON SAFETY • The warning signs and the sample icons shown here are listed to insure safe and accurate use. • The icons and meanings are as follows. Warning sign Indicates matters in which death or severe bodily damage may arise as a result of incorrect handling.
  • Page 4 NOTES ON SAFETY Unplug the AC adapter from the electric outlet if this unit is unused for an extended period of time. Unplug the AC adapter from the electric outlet when installing, removing, or cleaning the unit. Confirm readings with a stethoscope when an irregular pulse wave is displayed or when the measured value is questionable or erratic.
  • Page 5 NOTES ON SAFETY Do not use a cellular phone near this unit. Do not use this unit in a vehicle. Do not install the parts and/or instruments not specified for this unit. Do not use a broken power cord or AC adapter. Do not install or store this unit where it may come in contact with water or liquid medication.
  • Page 6: Troubleshooting

    (1) Remove the cuff from the patient’s arm. (2) Turn off the power of the unit and unplug the AC adapter from the electric outlet. (3) Display “Out of use” on this unit so that it cannot be used. (4) Contact Omron Healthcare’s Customer Service at 1-877-216-1336.
  • Page 7: Know Your Unit

    FEATURES OF THE PRODUCT OMRON IntelliSense ® Blood Pressure unit, Model HEM-907XL is developed to measure blood pressure and pulse rate accurately and simply in a doctor’s office, examination room, or patient bedside. • One-button operation Simply wrap the cuff and push the START Button. Blood pressure and pulse rate are automatically measured by the oscillometric method.
  • Page 8: Components Of The Product

    Model: HEM-907-CS19 Applicable arm circumference: 7" to 9" (17 to 22 cm) Instruction Manual (48H907N) INSTRUCTION MANUAL IntelliSense ® Blood Pressure Monitor HEM-907XL Model Table of Contents Be Sure to Read This Section ® Exemptions...2 Notes on Safety...3 Know Your Unit Features of the Product ...7...
  • Page 9: Features And Functions

    FEATURES AND FUNCTIONS Names of the Parts Main unit 1. Display (See Page 10 for detail) 2. HIDE (non-display) Button 3. DC jack 4. P-SET (pressure setting) Knob 5. MODE Selector 1. Display: Displays blood pressure and, pulse rate readings, and oscillation pulse level. 2.
  • Page 10 FEATURES AND FUNCTIONS Display Diastolic blood pressure PULSE Contents when the AVG Mode is selected AVG. : Mean value : First reading 2nd : Second reading : Third reading Systolic blood pressure mmHg mmHg HIDE Pulse rate Ready to Measure / Pulse synchronization Pulse level / Number of irregular pulses...
  • Page 11: Functions And Setting

    FEATURES AND FUNCTIONS Functions setting (1) Inflation level setting AUTO (Automatic setting): Can be used when the SINGLE, AVG, or MANU Mode is selected. The unit estimates the systolic blood pressure during inflation and inflates to a proper cuff pressure (approximately 30-40 mmHg above the patient’s systolic pressure).
  • Page 12 FEATURES AND FUNCTIONS List of Functions for each Mode Measurement Function Inflation level set- ting Non-display func- tion Manual deflation control (4) AVG Function setting You can set the number of measurements, the waiting time until the 1st measurement, and the measurement interval for the AVG Mode.
  • Page 13: Preparations Before Measurement

    Connect the air plug of the cuff to the air connector securely when connecting to the main unit. ® Blood Pressure Monitor HEM-907XL plays an important role of Enlarged figure of the circled part Cuff side...
  • Page 14 HOW TO APPLY THE CUFF 3. Place the right or left hand of the patient with the palm of hand facing upward. 4. Align the Artery Position Mark with the brachial artery. 5. Wrap the cuff snugly using both hands and securely fasten it with the Velcro™...
  • Page 15: How To Use The Power Source

    HOW TO USE THE POWER SOURCE CAREFULLY FOLLOW THESE INSTRUCTIONS. For connection to a power supply not in the U.S.A., use an attachment plug adapter of the proper configuration for the power outlet. The power unit (AC Adapter) is intended to be correctly oriented in a vertical or floor mount position.
  • Page 16: How To Use The Ac Adapter

    HOW TO USE THE POWER SOURCE How to use the AC adapter Do not use this unit in the presence of flammable gas, or anesthetics, or in a high pressure oxygen room or oxygen tent. Do not touch the AC adapter with wet hands. Connect the AC adapter to the DC jack of the main unit [1] and the electric outlet [2].
  • Page 17: Installation And Replacement Of Battery Pack

    HOW TO USE THE POWER SOURCE Installation and Replacement of Battery Pack If battery fluid gets into your eye or comes in contact with skin, wash the affected area with water repeatedly. Immediately consult a doctor for treatment. Do not use the battery pack for devices other than for this unit.
  • Page 18: How To Measure Blood Pressure

    HOW TO MEASURE BLOOD PRESSURE If cuff inflation does not stop, remove the cuff or pull out the air tube from the main unit. General advice • Patients should not touch the unit. List of Measurement Modes To measure only once To measure two times (or three times)
  • Page 19 HOW TO MEASURE BLOOD PRESSURE 1. Push the ON/OFF (power) Button to turn on the power. 2. Set the MODE Selector to “SINGLE”. 3. Set the P-SET (inflation level) Knob to “AUTO” or to the target pressure value. 4. Measure the patient’s arm size, and wrap appropriate cuff over the patient’s arm.
  • Page 20: How To Measure

    HOW TO MEASURE BLOOD PRESSURE 5. Push the START Button to start the measurement. • Do not push the START Button without wrapping the cuff. • If you want to stop measurement, push the STOP Button. The cuff will rapidly deflate.
  • Page 21 HOW TO MEASURE BLOOD PRESSURE 1. Push the ON/OFF (power) Button to turn on the power. 2. Set the MODE Selector to “AVG”. The factory-set values are set as follows: • Number of measurements: 2 • Waiting time until the 1st measurement: 0 sec. •...
  • Page 22 HOW TO MEASURE BLOOD PRESSURE 5. Push the START Button to start the measurement. After the pre-select waiting time, the unit takes the 1st measurement. After displaying the results of 1st measurement, subsequent measures occur automatically at the specified intervals. •...
  • Page 23: Manu Mode

    HOW TO MEASURE BLOOD PRESSURE 1. Push the ON/OFF (power) Button to turn on the power. 2. Set the MODE Selector to “MANU”. 3. Set the P-SET (inflation level setting) Knob to “AUTO” or the target pressure value. 4. Measure the patient’s arm size and wrap appropriate cuff over the patient’s arm.
  • Page 24 HOW TO MEASURE BLOOD PRESSURE 5. Place the stethoscope on the patient’s arm. 6. Push the START Button to start the measurement. • Do not push the START Button without wrapping the cuff. • Do not squeeze or press the cuff during the measurement.
  • Page 25: How To Check Pressure Accuracy

    Check result Accuracy of the monitor is validated to be ±3 mmHg or 2% of standard manometer reading. If your result shows a difference exceeding the tolerance, contact Omron Healthcare’s Cusomer Service at 1-877-216-1336. (IN CHECK MODE)
  • Page 26: How To Clean The Unit After Use

    HOW TO CLEAN THE UNIT AFTER USE When cleaning this unit, unplug the AC adapter from the electric outlet. After cleaning this unit, dry it well, before plugging the AC adapter in the electric outlet. HOW TO CLEAN THE UNIT AFTER USE 1.
  • Page 27: List Of Error Codes

    • Confirm that the cuff is wrapped correctly (refer to pages 13 and 14). • Ask the patient not to move during the measure- ment. • Check the patient for arrhythmia. • Contact Omron Healthcare’s Cusomer Service toll-free at 1-877-216-1336.
  • Page 28: Troubleshooting

    TROUBLESHOOTING If the unit malfunctions during use, please check the following: Trouble Is the cuff wrapped correctly? Is the patient moving during The unit inflates to inflation? abnormally high (low) pressure. Does this patient have arrhythmia? Check the patient’s condition. Is the patient moving during the measurement? The monitor cannot...
  • Page 29: Caution

    This warranty extends only to the original retail purchaser. To obtain warranty service contact Omron Healthcare’s Customer Service by calling 1-877-216-1336 for the address of the repair location and the return shipping and handling fee. Information for warranty service is avalable on our website at www.omronhealthcare.com...
  • Page 30 Please note that specifications may be changed without prior notice. Medical Equipment with respect to electric shock, fire and mechanical hazards only in accordance with UL 2601-1 and CAN/CSA C22.2 No. 601.1 OMRON Digital Automatic Blood Pressure Monitor HEM-907XL (HEM-907-Z2) Digital display Oscillometric method Pressure;...
  • Page 31 NOTE...
  • Page 32 For Customer Service Call Toll Free 1-877-216-1336 Made in Japan Manufactured by OMRON HEALTHCARE Distributed by: OMRON HEALTHCARE, INC. 1200 Lakeside Drive Bannockburn, Illinois 60015 www.omronhealthcare.com Copyright © 2009 OMRON HEALTHCARE, INC. 1628519-0E...