Omron HEM-907 Instruction Manual

Omron HEM-907 Instruction Manual

Digital automatic blood pressure monitor
Hide thumbs Also See for HEM-907:

Advertisement

INSTRUCTION MANUAL
OMRON
Digital Automatic Blood Pressure Monitor
Model
HEM-907
Thank you very much for purchasing the
OMRON Digital Automatic Blood Pressure
Monitor.
Requests from OMRON to the
operators and the persons
responsible for maintenance
Please read thoroughly the "Notes on
Safety" of this Instruction Manual before
using this unit so that you can use it safely
and correctly after sufficient understanding.
After reading this Instruction Manual, please
keep it near the unit all the times for future
reference.
The managerial responsibility of using and
maintaining the medical electronic instrument
must be borne by the users including hospi-
tals and clinics.
Do not let anyone use this unit other than the
doctors or the legally qualified persons.
Furthermore, only the skilled persons can
use this unit.

Table of Contents

Exemptions ...................................................2
Notes on Safety ............................................7
Components of the Product ........................11
Options........................................................11
How to Apply the Arm Cuff .........................12
(AC Adapter) ...........................................13
SINGLE Mode.........................................15
AVG. Mode ..............................................16
MANU. Mode...........................................17
CHECK Mode .........................................18
Battery Pack............................................19
List of Error Codes......................................21
Troubleshooting...........................................22
Specifications..............................................23
1

Advertisement

Table of Contents
loading

Summary of Contents for Omron HEM-907

  • Page 1: Table Of Contents

    Digital Automatic Blood Pressure Monitor Model HEM-907 Thank you very much for purchasing the Table of Contents OMRON Digital Automatic Blood Pressure Exemptions ...........2 Monitor. Names and Functions of the Parts ....3 Notes on Safety ..........7 Requests from OMRON to the Components of the Product ......11...
  • Page 2: Exemptions

    1. When a problem or damage occurs caused by the maintenance and/or repair con- ducted by a person other than OMRON or the dealer specified by OMRON 2. The problem or damage of OMRON product caused by the product of other man- ufacturer not delivered by OMRON 3.
  • Page 3: Names And Functions Of The Parts

    Names and Functions of the Parts Main unit 1. LCD display 6. ON/OFF (power) Button 2. HIDE (non-display) Button 7. START Button 8. DEFLATION (deflation control) Button 3. DC jack 9. Arm Cuff Connector 4. P-SET (pressure setting) Volume 10. STOP Button 5.
  • Page 4 Names and Functions of the Parts Ready to Measure Display Pulse level / Number of irregular pulse waves Number of irregular pulse waves: The pulse wave having a difference of more than ±25% from the average Systolic blood interval of pulse waves is determined pressure as irregular.
  • Page 5 Names and Functions of the Parts Explanation of Functions (1) Pressure setting function P - SET Inflation can be set by AUTO (automatic setting) or the target value. AUTO (automatic setting): In the SINGLE (single measurement), AVG. (average), and MANU. (auscultation) Modes, the monitor estimates the systolic blood pressure value during inflation and automatically inflates to the proper value.
  • Page 6 Names and Functions of the Parts (4) Setting mode change function You can set the number of times to measure, the waiting time until the start of measurement, and the interval between measurements. Items to set Set value Number of measurements 2 times, 3 times Waiting time until the start of measurement 0 sec, 3 min, 5 min, 10 min.
  • Page 7: Notes On Safety

    Notes on Safety • The warning signs and the sample icons shown here are listed for you to use the product safely and correctly as well as to prevent the risk and the damage to you and others from happening. •...
  • Page 8 Notes on Safety Caution When you are not going to use the unit for a long period of time, be sure to remove the AC adapter from the electric outlet. • You may suffer electric shock, or electric leak or fire may arise because of deteriorated insulation. Be sure to remove the AC adapter from the electric outlet when installing, removing, or cleaning the part.
  • Page 9 Do not install or store the unit where it may be sprayed with water or medication. • You may suffer electric shock. Requests from OMRON • Do not place or put anything on the unit. • Do not apply strong shock to or drop the unit.
  • Page 10 2. Turn off the power and pull the AC adapter from the electric outlet. If an optional battery pack is used, remove it also. 3. Display "Trouble" on the unit so that it cannot be used. 4. Contact the dealer where you purchased the unit or the nearest OMRON dealer...
  • Page 11: Components Of The Product

    Components of the Product Main unit Accessories Instruction Manual Medium size arm cuff (with built-in air bag) (with guarantee card) Model: HEM-9CMC Art.no.: 4928640-0 Applicable arm circumference: 22 to 32 cm Air tube (1 m) AC adapter Battery pack Model: HEM-9ADAP Model: HEM-9BAT Art.no.: 4928653-2 Art.no.: 4928652-4...
  • Page 12: How To Apply The Arm Cuff

    How to Apply the Arm Cuff Warning Requests from OMRON • Do not use the unit to the patient using a pump oxy- Do not apply the arm cuff during the intravenous genator. injection or transfusion. • Do not inflate the unit without applying the arm cuff.
  • Page 13: How To Use The Power Source (Ac Adapter)

    Do not install or store the unit where it may be sprayed with water or medication. • You may suffer electric shock. Requests from OMRON • If this unit is used concurrently with other unit, be sure to • When using an optional battery pack, the AC adapter read the Instruction Manual of the other unit to be used functions as the charger also.
  • Page 14: How To Measure Blood Pressure

    · An error may be displayed when an external vibra- tion is applied or the patient moves his/her body. (2) When an error is generated or the measured Requests from OMRON value is doubtful · An error may be displayed when an external vibra- tion is applied or the patient moves his/her body.
  • Page 15 How to Measure Blood Pressure (in SINGLE Mode) Wrap the arm cuff over the Push the ON/OFF (power) patient's arm. Button to turn on the power. ON/OFF START HIDE P-SET MODE SINGLE MANU. AVG. CHECK STOP AUTO Set the MODE Selector to Set the P-SET (pressure set- "SINGLE".
  • Page 16 How to Measure Blood Pressure (in AVG. Mode) Wrap the arm cuff over the Push the ON/OFF (power) patient's arm. Button to turn on the power. ON/OFF START HIDE P-SET MODE SINGLE AVG. MANU. CHECK STOP AUTO Set the MODE Selector to Set the P-SET (pressure set- "AVG.".
  • Page 17 How to Measure Blood Pressure (in MANU. Mode) Wrap the arm cuff over the Place the stethoscope on the patient's arm. patient's arm. ON/OFF START HIDE P-SET MODE SINGLE AVG. MANU. CHECK STOP AUTO Push the ON/OFF (power) Set the MODE Selector to Button to turn on the power.
  • Page 18 (1) In the measuring rule, the test tolerance of blood pressure monitor (accuracy of pressure when shipped from the facto- ry) should not exceed 3 mmHg. (2) If the CHECK result is +/- 3 mmHg or above, contact the dealer where you purchased this unit or the nearest OMRON dealer.
  • Page 19: Installation And Replacement Of Battery Pack

    Installation and Replacement of Battery Pack Warning Caution Do not disassemble or modify the battery pack. Do not short the polarities of battery using metal Do not throw the battery pack into fire or heat it. object such as the wire. Do not charge the battery pack by using equip- If the fluid in the battery is stained on your skin ment other than this unit.
  • Page 20: How To Clean The Unit After Use

    How to Clean the Unit after Use Caution Requests from OMRON Pull the AC adapter from the electric outlet when • Do not clean the unit with gasoline, thinner, or cleaning the unit. high concentration alcohol. • You may suffer electric shock.
  • Page 21: List Of Error Codes

    • Pule rate did not stay within the range of 30 to 199 beats/min. Device error • Contact the dealer where you purchased the unit or • There is a problem with the main unit of the the nearest OMRON dealer. blood pressure monitor.
  • Page 22: Troubleshooting

    Omron subsidiary Omron Niederlassung Succursale Omron Healthcare UK LTD. Omron Medizintechnik OMRON SANTÉ FRANCE Opal Drive Handelsgesellschaft m.b.H. 14, rue de Lisbonne Fox Milne, Milton Keynes MK15 0DG Windeckstraße 81a...
  • Page 23: Specifications

    Specifications Name OMRON Digital Automatic Blood Pressure Monitor Model HEM-907 Display Digital display Measurement Oscillometric method Measurement Range Pressure; 0 to 299 mmHg Pulse rate: 30 to 199 beats/min Accuracy Pressure; Within +/-3 mmHg Pulse rate; Within +/-5% of reading...
  • Page 24 Important information for users IM-HEM907E-02-10/06...