Omron MIT3 - Automatic Blood Pressure Monitor Manual

Omron MIT3 - Automatic Blood Pressure Monitor Manual

Also See for MIT3:

Advertisement

Omron MIT3 - Automatic Blood Pressure Monitor Manual

Introduction

Thank you for purchasing the OMRON MIT3 Automatic Blood Pressure Monitor.
Your new blood pressure monitor uses the oscillometric method of blood pressure measurement. This means the monitor detects your blood movement through your brachial artery and converts the movements into a digital reading. An oscillometric monitor does not need a stethoscope so the monitor is simple to use.

Intended Use
This device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population. The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with the measurement result.

information Please read this instruction manual thoroughly before using the device.
Please keep for future reference. For specific information about your own blood pressure, CONSULT YOUR PHYSICIAN.

Important Safety Information


Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.

(General Usage)

  • DO NOT adjust medication based on measurement results from this blood pressure monitor. Take medication as prescribed by your physician. Only a physician is qualified to diagnose and treat High Blood Pressure.
  • The monitor is not intended to be a diagnostic device.
  • Consult your physician before using the device for any of the following conditions: common arrhythmias such as atrial or ventricular premature beats or atrial fibrillation, arterial sclerosis, poor perfusion, diabetes, pregnancy, pre-eclampsia, renal diseases.
    warning Note that PATIENT motion, trembling, shivering may affect the measurement reading.
  • Do not use the device on the injured arm or the arm under medical treatment.
  • Do not apply the arm cuff on the arm while on an intravenous drip or blood transfusion.
  • Consult your physician before using the device on the arm with an arterio-venous (A-V) shunt.
  • Do not use the device with other medical electrical (ME) equipment simultaneously.
  • Do not use the device in the area of HF surgical equipment, MRI, or CT scanner, or in an oxygen rich environment.
  • The air tube or the AC adapter cable may cause accidental strangulation in infants.
  • Contains small parts that may cause a choking hazard if swallowed by infants.

(AC Adapter (optional) Usage)

  • Do not use the AC adapter if the device or the power cord is damaged. Turn off the power and unplug the power cord immediately.
  • Plug the AC adapter into the appropriate voltage outlet. Do not use in a multi-outlet plug.
  • Never plug in or unplug the power cord from the electric outlet with wet hands.


Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or patient or damage to the equipment or other property.

(General Usage)

  • Always consult your physician. Self-diagnosis of measurement results and self-treatment are dangerous.
  • Consult your physician before using the device for any of the following conditions:
    • If you have had a mastectomy.
    • Do not take measurements more than necessary. It may cause bruising due to blood flow interference.
    • People with severe blood flow problems or blood disorders as cuff inflation can cause bruising.
  • Remove the arm cuff if it does not start deflating during the measurement.
  • Do not use this device on infants or persons who cannot express their intentions.
  • Do not use the device for any purpose other than measuring blood pressure.
  • Use only the approved arm cuff for this device. Use of other arm cuffs may result in incorrect measurement results.
  • Do not use a mobile phone or other devices that emit electromagnetic fields near the device. This may result in incorrect operation of the device.
  • Do not disassemble the monitor or arm cuff. This may cause an inaccurate reading.
  • Do not use in a location with moisture, or a location where water may splash on the device. This may damage the device.
  • Do not use the device in a moving vehicle (car, airplane).
  • Read "If your systolic pressure is more than 210 mmHg" of this instruction manual, if your systolic pressure is known to be more than 210 mmHg. Inflating to a higher pressure than necessary may result in bruising where the cuff is applied.

(AC Adapter (optional) Usage)

  • Fully insert the power plug into the outlet.
  • When disconnecting the power plug from the outlet, do not pull the power cord.
    Be sure to pull from the power plug safely.
  • When handling the power cord, take care not to do the following:
    • Do not damage.
    • Do not break it.
    • Do not tamper with it.
    • Do not forcibly bend or pull.
    • Do not twist.
    • Do not bundle during use.
    • Do not pinch.
    • Do not place under heavy objects.
  • Wipe the dust off from the power plug.
  • Unplug monitor when not in use.
  • Disconnect the power plug before cleaning.
  • Use only an OMRON AC adapter designed for this device. Use of unsupported adapters may damage and/or may be hazardous to the device.

(Battery Usage)

  • Do not insert the batteries with their polarities incorrectly aligned.
  • Use only 4 "AA" alkaline or manganese batteries with this device. Do not use other types of batteries. Do not use new and used batteries together.
  • Remove the batteries if the device will not be used for three months or more.

General Precautions

  • Do not forcibly crease the arm cuff or the air tube excessively.
  • Do not press the air tube while taking a measurement.
  • To unplug the air plug, pull on the air plug at the connection with the monitor, not the tube itself.
  • Do not drop the monitor or subject device to strong shocks or vibrations.
  • Do not inflate the arm cuff when it is not wrapped around your arm.
  • Do not use the device outside the specified environment. It may cause an inaccurate reading.
  • Read and follow the "Important information regarding Electro Magnetic Compatibility (EMC)" in the "6. Specifications".
  • Read and follow the "Correct Disposal of This Product" in "6. Specifications" when disposing of the device and any used accessories or optional parts.

Know Your Device

Contents

Monitor, arm cuff, instruction manual, storage case, battery set

Monitor

Monitor

  1. Display
  2. Memory button
  3. START/STOP button
  4. Up/Down buttons
  5. Date/Time setting button
  6. Battery compartment
  7. AC adapter jack (for optional AC adapter)
  8. Air jack

Arm cuff

Arm cuff

  1. Arm cuff (Arm circumference 22 - 42 cm)
  2. Air plug
  3. Air tube

Display

Display

  1. Systolic blood pressure
  2. Diastolic blood pressure
  3. Low battery symbol
  4. Cuff wrap guide symbol
  5. Heartbeat symbol
    1. Flashes during measurement.
    2. If flashing after measurement complete indicates blood pressure out of recommended range.
  6. Memory symbol
  7. Date/Time display
  8. Irregular heartbeat symbol
  9. Blood pressure level indicator
  10. Deflation symbol
  11. Pulse display / Memory number

Display symbols

Irregular Heartbeat Symbol

Normal Heartbeat
Normal Heartbeat

Irregular Heartbeat
Irregular Heartbeat

When the monitor detects an irregular rhythm two or more times during the measurement, the irregular heartbeat symbol will appear on the display with the measurement values.

An irregular heartbeat rhythm is defined as a rhythm that is 25% less or 25% more than the average rhythm detected while the monitor is measuring the systolic and diastolic blood pressure.

If the irregular heartbeat symbol displays with your measurement results, we recommend you consult your physician. Follow the directions of your physician.

Cuff Wrap Guide Symbol

If the cuff was wrapped too loosely, it may cause unreliable values. If the wrapping of cuff is too loose, the cuff wrap guide symbol appears.
Otherwise appears. This is the function which is used as an aid in determining if the cuff is wrapped snugly enough.

Blood pressure Level Indicator

Blood pressure Level Indicator
Blood pressure level indicator will light up between the systolic blood pressure and the diastolic blood pressure. If your systolic or diastolic pressure is above the standard range (135 mmHg for the systolic blood pressure and/or 85 mmHg for the diastolic blood pressure), the heartbeat symbol will flash when the measurement result is displayed.

2013 ESH/ESC Guidelines for the management of arterial hypertension
Recent research suggests that the following values can be used as a guide to high blood pressure for measurements taken in office or at home.

Office Home
Systolic Blood Pressure ≥ 140 mmHg ≥ 135 mmHg
Diastolic Blood Pressure ≥ 90 mmHg ≥ 85 mmHg

Before Taking a Measurement

To help ensure an accurate reading, follow these directions:

  1. Avoid bathing, drinking alcohol or caffeine, smoking, exercising and eating for 30 minutes before taking a measurement. Rest for at least 15 minutes before taking the measurement.
  2. Stress raises blood pressure. Avoid taking measurements during stressful times.
  3. Measurements should be taken in a quiet place.
  4. Remove tight-fitting clothing from your arm.
  5. Sit on a chair with your feet flat on the floor. Rest your arm on a table so that the arm cuff is at the same level as your heart.
  6. Remain still and do not talk during the measurement.
  7. Keep a record of your blood pressure and pulse readings for your physician. A single measurement does not provide an accurate indication of your true blood pressure. You need to take and record several readings over a period of time. Try to measure your blood pressure at the same time each day for consistency.

Preparation

Battery Installation

Battery Installation

  1. Remove the battery cover.
  2. Insert 4 "AA" batteries as indicated in the battery compartment.
  3. Replace the battery cover.

warning Notes:

  • When the low battery symbol ( ) appears on the display, turn the monitor off, then replace all batteries at the same time. Long life alkaline batteries are recommended.
  • The measurement values continue to be stored in memory even after the batteries are replaced.
  • The supplied batteries may have a shorter life.
  • Dispose of the device, components and optional accessories according to applicable local regulations. Unlawful disposal may cause environmental pollution.

Setting the Date and Time

Set the monitor to the correct date and time before taking a measurement for the first time.

  1. Press the button.
  2. Push or to change the year
    Push button to confirm the year and then the month flashes. Repeat the same steps to change the month, day, hour, and minute.
    Setting the Date and Time
    : Forward
    : Back
  3. Press the START/STOP button to turn the monitor off.

warning Notes:

  • If the batteries have been replaced, the date and time setting will need to be reset.
  • If the date and time are not set, "-:--" appears during or after measurement.

Using the Device

Applying the Arm Cuff

Remove tight-fitting clothing or tight rolled up sleeve from your left upper arm.
Do not place the arm cuff over thick clothes.

  1. Insert the air plug into the air jack securely.
    Applying the Arm Cuff - Step 1
  2. Wrap the arm cuff firmly in place around your left upper arm.

    The bottom edge of the arm cuff should be 1 to 2 cm above the elbow. Air tube is on the inside of your arm and aligned with your middle finger.
    Applying the Arm Cuff - Step 2
  3. Secure closed with the fabric fastener.
    Applying the Arm Cuff - Step 3

warning Notes:

  • When you take a measurement on the right arm, the air tube will be at the side of your elbow. Be careful not to rest your arm on the air tube.
  • The blood pressure can differ between the right arm and the left arm, and the measured blood pressure values can be different. OMRON recommends to always use the same arm for measurement. If the values between both arms differ substantially, please check with your physician which arm to use for your measurements.

How to Sit Correctly

To take a measurement, you need to be relaxed and comfortably seated, under comfortable room temperature. Avoid bathing, drinking alcohol or caffeine, smoking, exercising or eating 30 minutes before taking a measurement.
How to Sit Correctly

  • Sit on a chair with your feet flat on the floor.
  • Sit upright with your back straight.
  • Sit with your back and arm being supported.
  • The arm cuff should be placed on your arm at the same level as your heart.

Taking a Measurement

warning Notes:

  • To stop a measurement, press the START/STOP button once to release the air in the arm cuff.
  • Remain still and quiet while taking a measurement.
  1. Press the START/STOP button.
    The arm cuff will start to inflate automatically.

Taking a Measurement
If your systolic pressure is more than 210 mmHg
After the arm cuff starts to inflate, press and hold the START/STOP button until the monitor inflates 30 to 40 mmHg higher than your expected systolic pressure.
warningNotes:

  • The monitor will not inflate above 299 mmHg.
    warning Inflating to a higher pressure than necessary may result in bruising where the arm cuff is applied.
  1. Remove the arm cuff.
  2. Press the START/STOP button to turn the monitor off.
    The monitor automatically stores the measurement result in its memory. It will automatically turn off after 2 minutes without any operations.
    warning Note: Wait 2-3 minutes before taking another measurement. Waiting between measurements allows the arteries to return to the condition prior to taking a measurement.

Using the Guest Mode
No measurement values are stored in the memory when the guest mode is selected.

  1. Press and hold the START/STOP button for more than 3 seconds.
    The Date/Time display will disappear.
    Using the Guest Mode
  2. Release the START/STOP button when the Date/Time display turns off.
    The arm cuff will start to inflate automatically.
  • DO NOT adjust medication based on measurement results from this blood pressure monitor. Take medication as prescribed by your physician. Only a physician is qualified to diagnose and treat High Blood Pressure.
  • This monitor is not intended to be a diagnostic device.
  • Always consult your physician. Self-diagnosis of measurement results and self-treatment are dangerous.
  • Read "If your systolic pressure is more than 210 mmHg" of this instruction manual, if your systolic pressure is known to be more than 210 mmHg. Inflating to a higher pressure than necessary may result in bruising where the cuff is applied.

Using the Memory Function

The monitor automatically stores the results up to 60 sets.

warning Notes:

  • If the memory is full, the monitor will delete the oldest value.
  • When viewing the measurement value taken without setting the date and time, "-:--" is displayed instead of the date and time.

To View the Measurement Values Stored in Memory

  1. Press the button.
    The Memory number appears for a second before the pulse rate is displayed. The newest set is numbered "1".
    To View the Measurement Values Stored in Memory
    warning Note: The cuff wrap guide result appears on the display with the measurement values.
  2. Press the or button to view the values stored in memory.
    : To view the older values
    : To view the more recent values

To Delete All the Values Stored in Memory

  1. Press the Memory button, while the memory symbol () appears.
  2. While holding the button down, press the START/ STOP button for more than 3 seconds.
    To Delete All the Values Stored in Memory

warning Note: You cannot partially delete the values stored in the memory.
All values will be deleted.

Error Messages and Troubleshooting

Error Messages

Error Display Cause Solution
Irregular heartbeats are detected. Remove the arm cuff. Wait 2 - 3 minutes and then take another measurement. Repeat the steps in section "Taking a Measurement". If this error continues to appear, contact your physician.
Arm cuff is applied too loosely. Apply the arm cuff tighter. Refer to "Applying the Arm Cuff".
The batteries are low. You should replace the batteries with new ones ahead of time.
Refer to "Battery Installation".
The batteries are exhausted. You should replace the batteries with new ones at once.
Refer to "Battery Installation".
Air plug disconnected. Insert the plug securely. Refer to "Applying the Arm Cuff".
Arm cuff is applied too loosely. Apply the arm cuff tighter. Refer to "Applying the Arm Cuff".
Air is leaking from the arm cuff. Replace the cuff with a new one. Refer to "Optional Medical Accessories".
Movement during measurement and the arm cuff has not been inflated sufficiently. Repeat measurement. Remain still and do not talk during measurement.
Refer to "Taking a Measurement".
If "E2" appears repeatedly, inflate the cuff manually until it is 30 to 40 mmHg above your previous measurement result.
Refer to "Taking a Measurement".
The arm cuff was inflated exceeding the maximum allowable pressure, and then deflated automatically when inflating the arm cuff manually. Do not touch the arm cuff and/or bend air tube while taking a measurement. Do not inflate the arm cuff more than necessary.
Refer to "Taking a Measurement".
Movement during measurement. Repeat measurement. Remain still and do not talk during measurement.
Refer to "Taking a Measurement".
Clothing is interfering with the arm cuff. Remove any clothing interfering with the arm cuff.
Refer to "Applying the Arm Cuff".
Device error. Contact your OMRON retail outlet or distributor.

Troubleshooting

Problem Cause Solution
The measurement result is extremely high (or low). Arm cuff is applied too loosely. Apply the arm cuff tighter. Refer to "Applying the Arm Cuff".
Movement or talking during measurement. Remain still and do not talk during measurement.
Refer to "Taking a Measurement".
Clothing is interfering with the arm cuff. Remove any clothing interfering with the arm cuff.
Refer to "Applying the Arm Cuff".
Arm cuff pressure does not rise. The air connector is not securely connected into the air jack. Make sure that the air tube is connected securely.
Refer to "Applying the Arm Cuff".
Air is leaking from the arm cuff. Replace the arm cuff with a new one.
Refer to "Optional Medical Accessories".
Arm cuff deflates too soon. The arm cuff is loose. Apply the cuff correctly so that it is firmly wrapped around the arm.
Refer to "Applying the Arm Cuff".
Cannot measure or the results are too low or too high. The arm cuff has not been inflated sufficiently. Inflate the cuff so that it is 30 to 40 mmHg above your previous measurement result.
Refer to "Taking a Measurement".
Nothing happens when you press the buttons. The batteries are empty. Replace the batteries with new ones.
Refer to "Battery Installation".
The batteries have been inserted incorrectly. Insert the batteries with the correct (+/-) polarity.
Refer to "Battery Installation".
Other problems.
  • Press the START/STOP button and repeat measurement.
  • Replace the batteries with new ones.
If the problem continues, contact your OMRON retail outlet or distributor.

Maintenance and Storage

Maintenance

To protect your device from damage, please observe the following:

  • Store the device and the components in a clean, safe location.
  • Do not use any abrasive or volatile cleaners.
  • Do not wash the device and any components or immerse them in water.

  • Do not use petrol, thinners or similar solvents to clean the device.
  • Use a soft and dry cloth, or a soft and moistened cloth and neutral soap to clean on the monitor and the arm cuff.

  • Changes or modification not approved by the manufacturer will void the user warranty. Do not disassemble or attempt to repair the device or components. Consult your authorised OMRON retail outlet or distributor.

Calibration and Service

  • The accuracy of this blood pressure monitor has been carefully tested and is designed for a long service life.
  • It is generally recommended to have the device inspected every 2 years to ensure correct functioning and accuracy. Please consult your authorised OMRON retail outlet or distributor.

Storage

Keep the device in its storage case when not in use.

  1. Unplug the air plug from the air jack.
  2. Gently fold the air tube into the arm cuff.

    warning Note: Do not bend or crease the air tube excessively.
  3. Place the monitor and the arm cuff in the storage case.

Do not store the device in the following situations:

  • If the device is wet.
  • Locations exposed to extreme temperatures, humidity, direct sunlight, dust or corrosive vapors such as bleach.
  • Locations exposed to vibrations, shocks or where it will be at an angle.

Optional Medical Accessories

(within the scope of EC Medical Device Directive 93/42/EEC)

Arm cuff
Arm circumference
22 - 42 cm
Small arm cuff
Arm circumference
17 - 22 cm
AC adapter
Easy Cuff (HEM-RML31-E) CS2 (HEM-CS24-E) AC ADAPTER-S (HEM-ACW5-E)

Using the Optional AC Adapter

Using the Optional AC Adapter

  1. Insert the AC adapter plug into the AC adapter jack on the rear side of the monitor.
  2. Plug the AC adapter into an electrical outlet.

To disconnect the AC adapter, unplug the AC adapter from the electrical outlet first, and then remove the AC adapter plug from the monitor.

Specifications

Product description OMRON Automatic Blood Pressure Monitor
Model MIT3 (HEM-7270-E)
Display LCD Digital Display
Measurement method Oscillometric method
Measurement range Pressure: 0 to 299 mmHg
Pulse: 40 to 180 beats/min.
Accuracy Pressure: ±3 mmHg
Pulse: ±5% of display reading
Inflation Fuzzy-logic controlled by electric pump
Deflation Automatic pressure release valve
Memory 60 measurements with date and time
Rating DC6V 4W
Power source 4 "AA" batteries 1.5V or optional AC adapter
(AC ADAPTER-S, INPUT AC100-240V 50/60Hz 0.12A)
Battery life Approx. 1000 measurements (using new alkaline batteries, operating temperature and humidity 23°C, 65% RH, cuff circumference 25 cm, maximum pressure 170 mmHg)
Applied part =Type BF
Protection against electric shock Internally powered ME equipment (When using only the batteries)
= Class II ME equipment (Optional AC adapter)
Operating temperature/ humidity/ air pressure +10°C to +40°C / 15% to 85% RH / 700 to 1060 hPa
Storage temperature/ humidity/ air pressure -20°C to +60°C / 10% to 95% RH / 700 to 1060 hPa
IP classification IP 20
Weight Monitor: Approx. 330 g without batteries
Arm cuff: Approx. 170 g
Outer dimensions Monitor: Approx. 115 (w) mm x 86 (h) mm x 145 (l) mm
Arm cuff: Approx. 145 mm x 594 mm
Cuff circumference 22 to 42 cm
Cuff/ Tube material Nylon, polyester, polyvinyl chloride
Package contents Monitor, arm cuff, instruction manual, storage case, battery set

warning Notes:

  • These specifications are subject to change without notice.
  • In the clinical validation study, the 5th phase was used on 85 subjects for determination of diastolic blood pressure.
  • This device has not been validated for use on pregnant patients.

Documents / Resources

Download manual

Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.

Download Omron MIT3 - Automatic Blood Pressure Monitor Manual

Advertisement

Need Assistance?

Do you have a question about the MIT3 that isn't answered in the manual? Leave your question here.

Error while submitting your question. Please, reload page and try again.

Thank you! Your question has been received!

Table of Contents