INSTRUCTION MANUAL
Blood Pressure Monitor
Model HEM-9200T
5130713-0C

Advertisement

Table of Contents
loading

Summary of Contents for Omron HEM-9200T

  • Page 1 ® INSTRUCTION MANUAL Blood Pressure Monitor Model HEM-9200T 5130713-0C...
  • Page 2: Table Of Contents

    Table of Contents Introduction ....................3 Intended Use ....................4 Important Safety Information ................5 1. Know Your Device .................8 2. Preparation ..................12 3. Using the Device..................15 4. Error Messages and Troubleshooting..........22 5. Maintenance and Storage..............27 6. Specifications..................30 7. FCC/IC Statement and Trademarks ............32 8.
  • Page 3: Introduction

    Introduction Please read this instruction manual thoroughly before using the device. Please keep for future reference. For specific information about your own blood pressure, CONSULT YOUR PHYSICIAN. Symbols Glossary For symbol information, visit: OmronHealthcare.com/symbols-glossary...
  • Page 4: Intended Use

    Intended Use The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population. The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings.
  • Page 5: Important Safety Information

    Important Safety Information Warning: Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury. General Usage • DO NOT adjust medication based on measurement results from this blood pressure monitor. Take medication as prescribed by your physician. Only a physician is qualified to diagnose and treat High Blood Pressure.
  • Page 6 Data Transmission • Do not use this product on aircraft or in hospitals. Please remove the battery from the unit. This product emits radio frequencies (RF) in the 2.4 GHz band, use of this product in locations where RF is restricted is not recommended. The use of RF in this product is licensed for use by the FCC, for further information ®...
  • Page 7 • Unplug monitor when not in use. • Disconnect the power plug before cleaning. • Use only an OMRON AC adapter designed for this device. Use of unsupported adapters may damage and/or may be hazardous to the device. Battery Usage •...
  • Page 8: Know Your Device

    1. Know Your Device Contents: Monitor, arm cuff, 4 “AA” batteries, instruction manual, quick start guide Monitor: A. Display C. Battery compartment B. START/STOP button D. Air jack E. AC adapter jack (for optional AC adapter) Arm cuff: Arm circumference 9" - 17" (22 - 42 cm) F.
  • Page 9 Display: SYNC symbol P. Movement error symbol J. Connection symbol Q. Irregular heartbeat symbol K. OK symbol R. Deflation symbol L. Systolic blood pressure S. Pulse display M. Diastolic blood pressure N. Battery symbol (low/depleted) O. Heartbeat symbol (Flashes during measurement.)
  • Page 10 1.1 Display symbols Irregular Heartbeat Symbol ( When the monitor detects an irregular Normal Heartbeat rhythm two or more times during the Pulse measurement, the irregular heartbeat symbol will appear on the display with the Blood pressure measurement values. Irregular Heartbeat An irregular heartbeat rhythm is defined as Short Long a rhythm that is 25% less or 25% more...
  • Page 11 Bluetooth Smart compatible devices and Telehealth service receiver. “Omron Healthcare, Inc. shall not be liable for loss of use or any other special, incidental, consequential or indirect costs, expenses or damages due to improper usage of this monitor.”...
  • Page 12: Preparation

    2. Preparation Battery Installation Below procedures are ONLY done at inserting batteries. 1. Remove the battery cover. 2. Insert 4 “AA” batteries as indicated in the battery compartment. 3. Put back the battery cover. Notes: • When the depleted battery symbol ( ) appears on the display, turn the monitor off, then replace all batteries at the same time.
  • Page 13 4. Connect the device to the Telehealth service receiver. As soon as insert batteries, it will automatically start the connection process, as below. When the display shown above is not appeared, see “Connecting failure.” in section 4.2 first. To retry connecting the Telehealth service receiver, remove batteries and press START/STOP button for 2-3 times.
  • Page 14 5. Confirm if the device is successfully connected. If the device is connected successfully to the Telehealth service receiver, OK symbol will appear on the display, as below. When “Err” appears, see “Connecting failure.” in section 4.2 for more detail. Notes: •...
  • Page 15: Using The Device

    3. Using the Device 3.1 Applying the Arm Cuff Remove tight-fitting clothing or tight rolled up sleeve from your left upper arm. Don’t place the arm cuff over thick clothes. 1. Connect cuff plug to the unit.
  • Page 16 2. Wrap the arm cuff firmly in place around your left upper arm. The bottom edge of the arm cuff should be 1/2 inch (1 to 2 cm) above the elbow. Air tube is on the inside of 1/2 inch your arm and aligned with your (1 - 2 cm) middle finger.
  • Page 17 • The blood pressure can differ between the right arm and the left arm, and the measured blood pressure values can be different. OMRON recommends to always use the same arm for measurement. If the values between both arms differ substantially, please check with your physician which arm to use for your measurements.
  • Page 18 3.2 How to Sit Correctly To take a measurement, you need to be relaxed and comfortably seated, under comfortable room temperature. • Sit on a chair with your legs uncrossed and your feet flat on the floor. • Sit upright with your back straight. •...
  • Page 19 1. Press the START/STOP button. The arm cuff will start to inflate automatically. START INFLATE DEFLATE COMPLETE What to do if your systolic pressure is more than 210 mmHg After the arm cuff starts to inflate, press and hold the START/ STOP button until the monitor inflates 30 to 40 mmHg higher than your expected systolic pressure.
  • Page 20 2. Transferring measured data. As soon as the measurement is completed, the data transfer will start automatically, as below. When the connection symbol ( ) is not appeared, see “Data is not being transmitted.” in section 4.2 for more detail. 3.
  • Page 21 4. Remove the arm cuff. DO NOT adjust medication based on measurement results from this blood pressure monitor. Take medication as prescribed by your physician. Only a physician is qualified to diagnose and treat High Blood Pressure. This device is not intended to be a diagnostic device. Always consult your physician.
  • Page 22: Error Messages And Troubleshooting

    4. Error Messages and Troubleshooting 4.1 Error Messages Error Display Cause Solution Remove the arm cuff. Wait 2 - 3 minutes and then take another Irregular heartbeats are measurement. Repeat the steps detected. in section 3.3. If this error continues to appear, contact your physician.
  • Page 23 Error Display Cause Solution Insert the plug securely. Air plug disconnected. Refer to section 3.1. Arm cuff is applied too Apply the arm cuff tighter. loosely. Refer to section 3.1. Replace the arm cuff with a Air is leaking from the new one.
  • Page 24 Error Display Cause Solution Repeat measurement. Remain Movement during still and do not talk during measurement. measurement. Refer to section 3.3. Remove any clothing interfering Clothing is interfering with the with the arm cuff. arm cuff. Refer to section 3.1. Contact your Telehealth service Device error.
  • Page 25 4.2 Troubleshooting Problem Cause and Solution Replace all batteries with new ones. No power. Check the battery installation for proper placement of the battery polarities. No display appears on the monitor. Refer to section 2. Blood pressure varies constantly. Many factors including stress, time of day, and Measurement values appear too high or how you wrap the cuff, may affect your...
  • Page 26 Problem Cause and Solution The blood pressure monitor might not be properly placed within the receiver’s transmission range and is too far from the receiver. If there are no causes of data transmission interference found near the blood pressure monitor, move the blood pressure monitor within 16 ft.
  • Page 27: Maintenance And Storage

    5. Maintenance and Storage 5.1 Maintenance To protect your device from damage, please follow the directions below: • Store the device and the components in a clean, safe location. • Do not use any abrasive or volatile cleaners. • Do not wash the device and any components by immersing them in water.
  • Page 28 5.2 Storage 1. Unplug the air plug from the air jack. 2. Gently fold the air tube into the arm cuff. Note: Do not bend or crease the air tube excessively. Do not store the device in the following situations: •...
  • Page 29 5.3 Optional Accessories Arm cuff AC Adapter Arm circumference Arm circumference 9" - 17" (22 - 42 cm) 7" - 9" (17 - 22 cm) CD-WR17 CD-CS9 HEM-ADPTW5 (Model: HEM-RML31) (Model: HEM-CS24) • Same as the arm cuff provided with the product.
  • Page 30: Specifications

    6. Specifications HEM-9200T Model Display LCD digital display Measurement range Pressure: 0 to 299 mmHg Pulse: 40 to 180 beats / min. Pressure: ±3 mmHg or 2% of reading Accuracy Pulse: ± 5% of display reading Inflation Fuzzy-logic controlled by electric pump...
  • Page 31 Notes: • These specifications are subject to change without notice. • In the clinical validation study, the 5th phase was used on 85 subjects for determination of diastolic blood pressure. • This device has not been validated for use on pregnant patients. •...
  • Page 32: Fcc/Ic Statement And Trademarks

    7. FCC/IC Statement and Trademarks FCC CAUTION Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment. This device complies with part 15 of FCC Rules and Innovation, Science and Economic Development Canada’s licence-exempt RSS(s).
  • Page 33 ® word mark and logos are registered trademarks The Bluetooth owned by Bluetooth SIG, Inc. and any use of such marks by OMRON HEALTHCARE Co., Ltd. is under license. Other trademarks and trade names are those of their respective owners.
  • Page 34: Limited Warranty

    2 years from the date of purchase, when used in accordance with the instructions provided with the monitor. The above warranty extends only to the original retail purchaser, and only to products purchased from an Omron authorized seller who is subject to and follows Omron’s quality control standards, unless otherwise prohibited by law.
  • Page 35: Guidance And Manufacturer's Declaration

    Nevertheless, special precautions need to be observed: • The use of accessories and cables other than those specified by OMRON, with the exception of cables sold by OMRON as replacement parts for internal components, may result in increased emission or decreased immunity of the device.
  • Page 36 Guidance and manufacturer’s declaration - electromagnetic emissions OMRON BPM including AC-adapter is intended for use in the electromagnetic environment specified below. The customer or the user of this OMRON BPM including AC-adapter should assure that it is used in such environment.
  • Page 37 Guidance and manufacturer’s declaration - electromagnetic immunity OMRON BPM including AC-adapter is intended for use in the electromagnetic environment specified below. The customer or the user of this OMRON BPM including AC-adapter should assure that it is used in such environment.
  • Page 38 Guidance and manufacturer’s declaration - electromagnetic immunity OMRON BPM including AC-adapter is intended for use in the electromagnetic environment specified below. The customer or the user of this OMRON BPM including AC-adapter should assure that it is used in such environment.
  • Page 39 OMRON BPM including AC-adapter OMRON BPM including AC-adapter is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of this OMRON BPM including AC-adapter can help prevent electromagnetic interference by maintaining a minimum...