Omron HEM-9000AI Instruction Manual page 89

Non-invasive blood pressure monitor with augmentation index (ai)
Hide thumbs Also See for HEM-9000AI:
Table of Contents

Advertisement

Error
Error Name
No.
Excessive pressure
Overpressure
was applied to the AI
E66
Error.
pulse wave sensor
unit.
AI pulse wave
Flash ROM
E67
measurement unit
Aberration.
malfunction.
Measurement
AI Pulse Wave
started while the AI
E70
Sensor is
pulse wave sensor
stored.
was still in the
storage clip.
AI Pulse Wave
Put the AI pulse
E71
Sensor is out of
wave sensor back in
storage.
the storage clip.
The pulse wave
No Pulse Wave
E72
cannot be accurately
Found.
detected.
The pulse wave was
Body
not accurately
E73
Movement
detected due to body
Error.
movement or
vibration.
Abnormal operation
is detected. Hold
Operation
E74
down pressure was
Aberration.
manually set outside
the range.
Measurement was
interrupted due to
Too Much Time
E75
time-out failure.
for
Measurement.
Measurement time
exceeded 5 minutes.
Communication
E76
Communication error
Error.
is detected between
the main unit and the
AI pulse wave
measurement unit.
Communication
E77
Error.
The AI pulse wave
Best Pressure
E80
sensor unit could not
Uncalculated.
detect a pulse.
Error Message on
Display
Measurement error Too
much pressure is on the
sensor.
The AI pulse wave
measurement unit has
failed.
Measurement began
while the AI pulse wave
sensor was stored.
The AI pulse wave
sensor has been out of
storage for an extensive
period of time.
The pulse wave is not
detectable.
The pulse wave was not
accurately detected due
to body movement.
Measurement Error /
Detect abnormal
operation.
Measurement Error /
Inflation time exceeds 5
minutes.
The AI pulse wave
measurement unit is not
communicating with the
main unit.
The AI pulse wave
measurement unit was
not communicating with
the main unit.
An irregular pulse wave
or similar problem
prevented the AI from
being measured.
Error Content
• There is a possibility that the AI
pulse wave sensor hit against
bone or something hard. After
reviewing Section 9.2., retake
the measurement.
• Turn off the power. Turn the
power back on. If the problem is
not solved, contact OMRON.
• Follow the instructions for
measurement preparation.
Refer to Section 9.
• While holding down the slide
stoppers on both sides, place
the AI pulse wave sensor in its
original position.
• There is a possibility that the AI
pulse wave sensor is not on the
artery. Reattach the AI pulse
wave sensor unit properly.
Refer to Section 9.2. Retake the
measurement.
• Check for arrhythmia.
• Make sure that the patient does
not move their arm or body
during measurement.
• After reviewing Section 14.,
retake the measurement.
• Reattach the AI pulse wave
sensor unit properly. Refer to
Section 9.2. Retake the
measurement.
• The main unit might not be
properly connected.
• After double-checking the cable
connection, retake the
measurement.
• Retake the measurement.
• There is a possibility that the AI
pulse wave sensor is not on the
artery. After reviewing Section
9.2., retake the measurement.
21.Error List
Action
EN
89