General Error Codes
Message Dialog
Main unit malfunction.
E10
Turn off main power supply switch.
End of Calendar.
Unable to continue.
E19
Change the date and time settings in the
settings menu after power has been
restored to the system.
Printer error.
E13
E20
Turn off main power supply switch.
No printer paper.
E21
Load paper.
Printer cover is open.
E22
Close cover.
Printer jam.
E23
Remove jammed paper.
This memory card has not been formatted.
E42
Push the eject button, remove the memory
card and insert a new memory card.
The measurement data has not been
E45
saved.
This system was not properly terminated.
E46
The latest measurement data has not been
saved.
There is no memory card inserted in the
device.
E47
Therefore the operation was terminated.
E48
ID DATA has not been copied.
Replace clock battery.
Insert memory card.
This memory card cannot be used.
When the memory card is not properly ejected,
important DATA may be lost. Turn the power off and
back on in order to insert a memory card.
Memory capacity is insufficient.
Replace with empty memory card.
Current memory card cannot be used.
This value is out of range.
If the errors cannot be solved by the suggested actions, contact OMRON.
Action
Turn off the power. Turn the power back on. If the
problem is not resolved, contact OMRON.
The calendar function cannot be used after
December 31, 2030.
Turn off the power. Turn the power back on. If the
problem is not resolved, contact OMRON.
Open the printer cover and load paper. (Refer to
Section 8.5.)
Close the printer cover.
Open the printer cover and remove the paper. (Refer
to Section 8.5.)
Obtain a new memory card and insert it into the
memory card slot on the main unit.
Retake the measurement.
Data was not saved since the power was
disconnected or interrupted during measurement.
Memory card has not been inserted. Insert a
memory card into the slot on the right side of the
main unit.
Try to copy the information again following the
procedure on Section 18.10. starting from the step
on copying ID information.
Contact OMRON.
Memory card has not been inserted. Insert a
memory card into the slot on the right side of the
main unit.
Format the memory card (Section 18.10.) or replace
it with a different memory card.
Turn off the power. Reinsert the memory card and
turn the power back on.
Eject the memory card and replace it with a new
one. (Section 18.10.)
Turn off the power. Reinsert the memory card and
turn the power back on.
The number just entered was not appropriate.
Reenter the number.