Note:
The difference in value of "Ref. Level" when sensing the mark should be greater than when sensing blank label
by no less than 128. If the sensor fails to differentiate by this much, manually increase Intensity until the above
difference is achieved.
BLine Mode
Complete
Intensity
Ref. Level
B-3 Pre-Printed
This function auto-calibrates the black line sensor more accurately for pre-printed labels.
BLine Mode
Manual
>
Pre Printed
Exit
Entering the Pre Printed option displays the following message, with steps which must be completed:
Mode
Paper Len.
00812 dot
BLine Size
0024 dot
BLIne Mode
Pre Printed
LabelStation Pro 240
X
XXX
3/4
Actions
1. Press the
right digit, and press the
and the "dot/ mm/ inch". Press the SELECT key to set the paper
length.
2. Press the
right digit, and press the
and the "dot/ mm/ inch". Press the SELECT key to set the gap
size.
3. The printer feeds labels to calibrate the sensor sensitivity
automatically. When calibration is complete, the LCD screen will
return to the previous menu.
Label Sensor
Blank label under sensor
5. BLine sensor calibration is complete. Press the SELECT
key to return the LCD screen to the previous menu.
DOWN
key
UP
DOWN
key
UP
Page 42 of 67
to move the cursor from left digit to
key
to set the value from "0" to "9"
to move the cursor from left digit to
key
to set the value from "0" to "9"