Checking And Replacing The Main Fuse - Invacare Spectra XTR Service Manual

Hide thumbs Also See for Spectra XTR:
Table of Contents

Advertisement

®
Invacare
- Bora
7.8

Checking and replacing the main fuse

CAUTION: Fire hazard!
A short circuit can cause extremely high currents which can result in spark formation and
fire!
 Always use an original strip fuse with the approved amperage.
 If the main fuse has blown, first rectify the cause before fitting a new one.
CAUTION: Fire hazard!
Fitting the incorrect strip fuse causes a fire hazard!
 Only fix the strip fuses in the sequence shown in the image on the right!
 Tighten the nuts with 3.3 to 3.5 Nm!
Requirements:
 Phillips screwdriver
 Strip fuse
 socket spanner, 8 mm
 torque wrench 0-20 Nm (or similar)
Note
A spare fuse is located behind the Invacare logo on the rear battery.
Note
If the fuseholder is damaged, you can replace this complete with the battery cables.
Remove both battery cases as
described in Chapter 7.7.1.
Remove the covers on both battery
cases as described in Chapter 7.7.2.
SERVICE MANUAL
1. Strip fuse
2. Spade terminal
3. DIN 6923 nut
59