VR2 remote: Error codes and diagnostic codes

Evaluate the cause using the following blink codes. The following figure shows which LEDs are
located on the remote.
1) Battery display
2) Profile indicator
3) Adjusting motors
Error code
1 LED battery display
2 LED battery display
3 LED battery display
4 LED battery display
5 LED battery display
6 LED battery display
7 LED battery display
8 LED battery display
POSSIBLE
SOLUTION
CAUSE
Batteries
Charge battery
discharged
Check cable to batteries
Error at left-hand
Check connection plug, check
motor
motor
Connection
loose/defective
or motor
defective
Short-circuit in
Check connection plug, check
left-hand motor
motor
Error at right-
Check connection plug, check
hand motor
motor
Connection
loose/defective
or motor
defective
Short-circuit in
Check connection plug, check
right-hand motor
motor
The mobility
Remove battery charger
device has been
blocked by an
external signal,
for example
because the
charger is
connected.
Fault on drive
Put the drive lever in a central
lever
position before switching the
remote on.
Fault in the
Check cable
electronics
SERVICE MANUAL
Documentation
See chapter 7.9.
See chapters 7.9
and 7.2
See chapters 7.9
and 7.2
See chapters 7.9
and 7.2
See chapters 7.9
and 7.2
See chapter 7.9.
31