Checking The Cable - Invacare Spectra XTR Service Manual

Hide thumbs Also See for Spectra XTR:
Table of Contents

Advertisement

®
Invacare
- Bora
7.9

Checking the cable

The following images show checking the cable using the Shark controller as an example. The plug
positions are different for both the DX2 and VR2 controllers. The procedure for checking is
otherwise identical.
The position of the plugs is described in Chapter 4.2.
Remove legrests.
Release seat (1) and tilt forwards.
Check all cables for visible damage and
crushing. Replace damaged cables.
Pull each plug (1) carefully. The plug
should not come out of the socket.
If the plug is loose, press the plug into
the socket with slight pressure. The plug
must engage.
Check whether the plug is now firmly
located in the socket, otherwise repeat
the previous step.
Remove both battery cases as described
in Chapter 7.7.1.
Remove the covers on both battery cases
as described in Chapter 7.7.2.
SERVICE MANUAL
61

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents