SERVICE MANUAL
8.4.5
Checking and replacing the main fuse (Standard TDX-SP)
CAUTION: Fire hazard!
A short circuit can cause extremely high currents which can result in spark formation and
fire!
· Always use an original strip fuse with the approved amperage.
· If the main fuse has blown, first rectify the cause before fitting a new one.
CAUTION: Fire hazard!
Fitting the incorrect strip fuse causes a fire hazard!
● Only fix the strip fuses in the sequence shown in the image on the right!
● Tighten the nuts with 3.3 or 3.5 Nm!
Requirements:
● 5/8" Allen key
● Phillips screwdriver size 2
● socket spanner, 8 mm
● torque wrench 0-20 Nm (or similar)
Note
If the fuseholder is damaged, you can replace this complete with the battery cables.
Dismantling the main fuse:
●
Remove the battery compartment cover, disconnect the battery plugs and pull the batteries out
of the battery compartment as described in Chapter 8.4.1.
●
The fuseholder (1) is located on top of the batteries.
●
Open the fuseholder cover.
58
®
Invacare
- TDX-SP / TDX-SP N
1. Strip fuse
2. Spade terminal
3. DIN 6923 nut