R-Net Remote: Error Codes And Diagnostic Codes - Invacare Spectra XTR Service Manual

Hide thumbs Also See for Spectra XTR:
Table of Contents

Advertisement

SERVICE MANUAL
6.5

R-NET remote: Error codes and diagnostic codes

Evaluate the cause using the following blink codes. The following figure shows which LEDs are
located on the remote.
1) Battery display
2) Actuators
3) Profile indicator
Error code
1 LED battery display
2 LED battery display
3 LED battery display
4 LED battery display
5 LED battery display
6 LED battery display
7 LED battery display
8 LED battery display
34
POSSIBLE
SOLUTION
CAUSE
Batteries
Charge battery
discharged
Check cable to batteries
Fault in left-hand
Check connection plug, check
motor;
motor
connection
loose/defective or
motor defective
Short-circuit in
Check connection plug, check
left-hand motor
motor
Fault in right-hand
Check connection plug, check
motor;
motor
connection
loose/defective or
motor defective
Short-circuit in
Check connection plug, check
right-hand motor
motor
The wheelchair
Remove battery charger
has been blocked
by an external
signal, for
example because
the charger is
connected.
Fault on drive
Put the drive lever in a central
lever
position before switching the
remote on.
Fault in the
Check cable
electronics
®
Invacare
- Bora
Documentation
See chapter 7.9.
See chapters 7.9
and 7.2
See chapters 7.9
and 7.2
See chapters 7.9
and 7.2
See chapters 7.9
and 7.2
See chapter 7.9.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents