Changing of circulating fluid discharge temperature drop setting and checking
8.
Press the [SEL] key once.
The set screen of changing of circulating fluid discharge temperature drop is displayed
on the digital display.
PV
SV
9.
Select changing of circulating fluid discharge temperature drop from the table below with
[▲] key or [▼] key, and confirm by pressing "SEL".
Set value
Detection temperature for circulating fluid discharge temperature drop Setting and checking
10.
Press the [SEL] key once.
The set screen of detection temperature for circulating fluid discharge temperature drop
is displayed on the digital display.
PV
SV
11.
Select detection temperature for circulating fluid discharge temperature drop from the
table below with [▲] key or [▼] key, and confirm by pressing "SEL".
Set value
Centigrade
to
Fahrenheit
to
HRS Series
Table 5.15-6 List of set value
Explanation
This alarm signal is not detected.
Operation continues when this alarm signal is
generated.
Operation is stopped when this alarm signal is
generated.
Table 5.15-7 List of set value
Explanation
Setting/checking are not available if the
setting of the circulating fluid discharge
temperature drop is OFF.
Sets detection temp for the circulating fluid
discharge temperature drop.
Temperature unit is Centigrade : Setting unit is
o
0.1
C
Temperature unit is Fahrenheit : Setting unit is
o
0.1
F
Chapter 5
Display and setting of various functions
HRX-OM-M090
Initial value
(Default setting)
Initial value
(Default setting)
5.15 Alarm customize function
5-39