Calibration - Cardinal PD150 Owner's Manual

Prodoc bfa series digital physician scale with body fat analysis
Table of Contents

Advertisement

Your ProDoc BFA Series Scale was calibrated at the factory and should not require adjustment.
In the event that the scale should need re-calibration, the following describes the calibration
procedure. To maintain the instrument's high degree of accuracy, a qualified technician should
perform this function.
Before beginning calibration, the following equipment is required:
150 kg (330 lbs) of calibrated test weight
To Enter Calibration
1. With the scale off, press and hold the ZERO and the MODE/ENTER/HEIGHT keys and
then press the ON/OFF key.
2. The display will turn on all digits, show the software version for a few seconds and then
change to show CAL.
3. Release all keys.
4. The display will change to show USA.
5. The scale is now ready for calibration.
IMPORTANT! When a prompt and value displayed are acceptable, press the
MODE/ENTER/HEIGHT key to save the setting and proceed to the next
prompt. To change a setting, use the numeric keys to enter a new value and
then press the MODE/ENTER/HEIGHT key to save the new setting and
advance to the next prompt.
USA (SET ZERO LIMIT)
This prompt selects whether the scale is used in the USA and has a zero limit or outside the
USA with no zero limit.
With display showing USA, press MODE/ENTER/HEIGHT key to show current setting. If
value displayed is acceptable, press MODE/ENTER/HEIGHT key to save it and proceed to
the next prompt. Otherwise, use the numeric keys to enter a new value 0 or 1 and then
press the MODE/ENTER/HEIGHT key to save it and proceed to the next prompt.
0 = +/- 2% limit in zero setting (with
maximum capacity up to +9d)
HoLd (HOLD MODE) "Lock Feature"
With display showing HOLD, press MODE/ENTER/HEIGHT key to show current setting. If
value displayed is acceptable, press MODE/ENTER/HEIGHT key to save it and proceed to
the next prompt. Otherwise, use the numeric keys to enter a new value 0 or 1 and then press
the MODE/ENTER/HEIGHT key to save it and proceed to the next prompt.
0 = Disable Hold Mode
NOTE: The Hold Mode "lock feature" is for non-commercial applications and must
be disabled for "Legal for Trade" applications.

CALIBRATION

1 = No limit in zero setting and
1 = Enable Hold Mode
0044-M380-O1 Rev A ProDoc BFA Series
maximum capacity up to +4%)
41