10. TOT CAP=XXXXXXX (TOTAL SCALE CAPACITY)
With SETUP MENU #2 displayed the current setting for the TOT CAP= parameter will be
shown. Note that XXXXXXX is the current value for the scale capacity. If the setting
displayed is acceptable, proceed to the next setup parameter.
Otherwise, press the 1 and 0 keys, the ENTER key and then using the numeric keys, enter
the value for the total scale capacity and then press the ENTER key to save the new setting
and return to SETUP MENU #2. Allowable values are 1 through 9,999,999. Note that the
total capacity can not exceed 7 digits or 9,999,999.
With the SETUP MENU #2 displayed, proceed to setup scale X for the setup of each scale.
IMPORTANT! All scales to be included in the totalizer must have identical settings
for: BASE UNITS, CNVT UNITS, INTERVAL, DDP, SAMPLES, STABLE
and STBL CNT. If these are not the same, an error message will be displayed.
For additional information on setup and calibration, refer to the 225 WEIGHT
INDICATOR Installation and Technical Manual, 8200-M538-O1.
With the SETUP/REVIEW MENU displayed, press the key corresponding to the scale to be
calibrated and then press the ENTER key. The display will change to show the SCALE X
CALIBRATION MENU, where X is the scale to be calibrated.
1.ENTER CALIBRATION AND SETUP
2.VIEW AUDIT TRAIL COUNTERS
3.CALIBRATE SCALE 1
4.CALIBRATE SCALE 2
5.CALIBRATE SCALE 3 10.PRINT SETUP
Enter Selection: 1
NOTE! Parameter 5 will only be shown if a 3 has been entered for the number of scales.
13
INDICATOR SETUP, CONT.

CALIBRATION

SETUP/REVIEW MENU
8200-M535-O1 Rev B  225-DS
9.DEL CUSTOM TICKET
^
EXIT