Calibration Procedure - Cardinal Detecto C Series Operation Manual

Digital counting scale
Table of Contents

Advertisement

To begin the calibration procedure, first remove the seal and the cover on the bottom of the
scale. Next, locate the test pins and short the right side 2 pins to enter the calibration
setting mode. The display will show SETUP, then SCALE2, nO.
To enable the use of the remote scale, press the TARE key to toggle the selection to YES.
With the appropriate selection made, press the UNITS key seven (7) times until the display
shows SETUP, CAL , the current setting for the Calibration Units parameter.
CAL – Calibration Units
With SETUP, CAL , the current setting for the Calibration Units parameter will be shown.
If the setting displayed is acceptable, press the UNITS key to proceed to the next
parameter.
Otherwise, use the TARE key to toggle between the selections and then press the UNITS
key to proceed to the next parameter. The available settings are:
=
kg (Kilograms)
<G
=
Lb (Pounds)
Lb
CAP1 – Local Scale Capacity
With SETUP, CAP1 displayed, the current setting for the Scale Capacity will be shown. If
the setting displayed is correct, press the UNITS key to proceed to the next parameter.
Otherwise, use the TARE key to toggle between the selections and then press the UNITS
key to save it and proceed to the next parameter. The available settings are:
=
30Lb
=
65Lb
=
100Lb

CALIBRATION PROCEDURE

Scale capacity is 30 pounds
Scale capacity is 65 pounds
Scale capacity is 100 pounds
8526-M260-O1 Rev C  C Series Operation
17