Keithley 2002 User Manual page 54

Hide thumbs Also See for 2002:
Table of Contents

Advertisement

SENSOR
This parameter is used to select the temperature sensor.
4-WIRE-RTD or RTD: If using a 4-wire or 3-wire RTD
sensor, select 4-WIRE-RTD. If using a 2-wire RTD, select
RTD. Selecting an RTD sensor displays the options for RTD
standards. Five are configured for common RTDs and the
sixth allows you to enter your own RTD factors:
• PT100, D100, F100, PT385 and PT3916 — Selects the
following default parameters for the RTD standards:
Type
Standard Alpha
PT100
ITS-90
0.003850
D100
ITS-90
0.003920
F100
ITS-90
0.003900
PT385
IPTS-68
0.003850
PT3916
IPTS-68
0.003916
• USER — This selection allows you to change one or
more of the RTD factors listed in the above table. The
equation using the factors are shown in Figure 2-11.
SPRTD: Choosing type SPRTD takes you to the SPRTD
COEFFICIENTS menu, which allows you to program the
following parameters:
R-ZERO: RTD 0˚C resistance value
A4: set A4 coefficient
B4: set B4 coefficient
A7: set A7 coefficient
B7: set B7 coefficient
C7: set C7 coefficient
Coefficients
Subrange #1: 13.8033K - 273.16K
Subrange #2: 24.5561K - 273.16K
Subrange #3: 54.3584K - 273.16K
Subrange #4: 83.8058K - 273.16K
Subrange #5: 234.3156K - 302.9146K
Subrange #6: 273.15K - 1234.93K
Subrange #7: 273.15K - 933.473K
Subrange #8: 273.15K - 692.677K
Subrange #9: 273.15K - 505.078K
Subrange #10: 273.15K - 429.7485K
Subrange #11: 273.15K - 302.9146K
RZero
Beta
Delta
( at 0°C)
0.10863
1.49990
100.00000
0.10630
1.49710
100.00000
0.11000
1.49589
100.00000
0.11100
1.50700
100.00000
0.11600
1.50594
100.00000
The ITS-90 standard provides two reference equations for
Standard Platinum Resistance Thermometers covering the
temperature range 18.8033K to 1234.93K. A single SPRTD,
however, usually cannot be used to cover the entire range.
The temperature range is therefore broken up into several
subranges. These subranges depend on the calibration points
of the temperature scale and are based on the melting or tri-
ple points of various pure substances. For an exact list of the
elements needed and details on RTD calibration, refer to
NIST Technical Note 1265 "Guidelines For Realizing the
International Temperature Scale of 1990". In each subrange,
the calibration constants required for that range are listed.
An SPRTD as supplied from the manufacturer will come
with a certificate of calibration that lists the calibration
constants and the temperature range supported. In all cases
except subranges #4 and #7, translation of the supplied
coefficients to Model 2002 values will be required. In most
cases, this translation is done simply entering the A value
(for example) supplied with the SPRTD into the A4 position
(temperatures less than 0˚C) or A7 value (temperatures
above 0˚C) required by the Model 2002. The same procedure
is used for translating the B and, where applicable, C
constants.
The following table should be helpful in translating SPRTD
coefficients to Model 2002 SPRTD coefficients. The Model
2002 supports SPRTD temperatures between 83.805K and
933.473K. For any of the subranges below, only
temperatures within the range from 83.805K to 933.47K will
be measured.
RTD coefficients to Model 2002 coefficients
Not applicable
A2 to A4, B2 to B4
A3 to A4, B3 to B4
No substitution needed
A5 to A4, A5 to A7, B5 to B4
B5 to B7, Set C7 = 0. A5 value entered for both
A4 and A7; B5 value entered for both B4 and B7
Not applicable
No substitution needed
A8 to A7, B8 to B7, Set C7 = 0
A9 to A7, B9 to B7, Set C7 = 0
A10 to A7, B10 to B7, Set C7 = 0
A11 to A7, Set B7 = 0, C7 = 0
Front Panel Operation
2-37