Keithley 2002 User Manual page 181

Hide thumbs Also See for 2002:
Table of Contents

Advertisement

IEEE-488 Reference
Table 3-10 (Continued)
SENSe command summary
Command
:TEMPerature
AVERage
:TCONtrol <name>
:TCONtrol?
:COUNt <n>
:COUNt?
[:STATe] <b>
[:STATe]?
:AUTO <b>
[:STATe] <b>
[:STATe]?
:AUTO <b>
:AUTO ONCE
:AUTO?
:TRANsducer <name>
:TRANsducer?
:RTD
:TYPE <name>
:TYPE?
:ALPHa <NRf>
:ALPHa?
:BETA <NRf>
:BETA?
:DELTa <NRf>
:DELTa?
:RZERo <NRf>
:RZERo?
:SPRTD
:RZERo <NRf>
:A4 <NRf>
:B4 <NRf>
:A7 <NRf>
:B7 <NRf>
:C7 <NRf>
:TCouple
:TYPE <name>
:TYPE?
:RJUNctionX
:RSELect <name>
:RSELect?
:SIMulated <n>
:SIMulated?
:REAL
:TCOefficient <n>
:TCOefficient?
:OFFSet <n>
:OFFSet?
:ACQuire
3-58
Description
Path to configure and control the filter.
Select filter type: (MOVing or REPeat).
Query filter type.
Specify filter count (1 to 100).
Query filter count.
Enable or disable filter.
Query state of digital filter.
Enable or disable auto filter.
Enable or disable filter.
Query state of digital filter.
Enable or disable auto filter.
Enable and then disable auto filter.
Query auto filter.
Select transducer: (RTD, FRTD, TCouple, DTC, or INTer-
nal).
Query transducer.
Path to configure RTD measurements:
Select RTD parameters: (PT100, D100, F100, PT385,
PT3916, USER, or SPRTD).
Query RTD type.
Specify constant for USER type (0 to 0.01).
Query alpha.
Specify constant for USER type (0 to 1).
Query beta.
Specify constant for USER type (0 to 5).
Query delta.
Specify constant for USER type (0 to 1000).
Query rzero.
Set RTD zero value.
Set A4 coefficient.
Set B4 coefficient.
Set A7 coefficient.
Set B7 coefficient.
Set C7 coefficient.
Path to configure TC measurements:
Select thermocouple type: (J, T, K, E, R, S, B, or N).
Query TC type.
Path to configure reference junctions (X = 1 to 5).
Select reference type: (SIMulated or REAL).
Query reference type.
Specify simulated temperature.
Query simulated temperature.
Command path to specify "real" reference junction.
Specify TC
Query TC
Specify voltage offset at 0°C.
Query voltage offset.
Update reference temperature.
Default
parameter
SCPI
Ref.
3.18.11
(Note 3)
10
(Note 4)
(Note 5)
(Note 4)
(Note 5)
FRTD
3.18.17
3.18.18
PT100
0.00385
0.111
1.507
100
100
3.18.19
J
3.18.20
SIMulated
23°C
0.00002
0.05463

Advertisement

Table of Contents
loading

Table of Contents