Signal - Keithley 2002 User Manual

Hide thumbs Also See for 2002:
Table of Contents

Advertisement

Query
Description
Program fragment
3.23.8
Description
Program fragment
3.23.9
Parameters
Query
Description
Program fragment
Parameters
:TIMer?
:TIMer? DEFault
:TIMer? MINimum
:TIMer? MAXimum
These commands are used to set the interval for the scan layer and measure layer timers. Note
that the timer is in effect only if the timer is the selected control source. Also, note that the arm
layer (Arm Layer 1) does not use a timer.
PRINT #1, "output 16; :trig:tim .25; tim?"
PRINT #1, "enter 16"

:SIGNal

:ARM[:SEQuence[1]][:LAYer[1]]:SIGNal
:ARM[:SEQuence[1]]:LAYer2:SIGNal
:TRIGger[:SEQuence[1]]:SIGNal
These action commands are used to bypass the specified control source and also when you do
not wish to wait for the programmed event. Keep in mind that the instrument must be waiting
for the appropriate event when the command is sent. Otherwise, an error occurs and this com-
mand is ignored.
PRINT #1, "output 16; :trig:sign"
TCONfigure commands
:PROTocol <name>
:TRIGger[:SEQuence[1]]:TCONfigure:PROTocol <name>
<name> = ASYNchronous
SSYNchronous
:PROTocol? Query programmed Trigger Link protocol
This command is used to select the protocol for the measure layer Trigger Link. With ASYN-
chronous selected, separate trigger lines are used for input and output triggers. With SSYNchro-
nous selected, a single trigger line is used for both input and output triggers.
Note that the arm and scan layers only use the asynchronous Trigger Link mode and therefore,
do not have to be selected.
PRINT #1, "output 16; :trig:tcon:prot ssyn; prot?"
PRINT #1, "enter 16"
:DIRection <name>
:ARM[:SEQuence[1]][:LAYer[1]]:TCONfigure:DIRection <name>Control arm Source Bypass
:ARM[:SEQuence[1]]:LAYer2:TCONfigure:DIRection <name>Control scan Source Bypass
:TRIGger[:SEQuence[1]]:TCONfigure:DIRection <name>Control measure Source Bypass
<name> = SOURce
ACCeptor
Query programmed timer interval
Query *RST default timer interval
Query lowest allowable timer interval
Query largest allowable timer interval
Bypass arm control source
Bypass scan control source
Bypass measure control source
' Bypass measure control source
Asynchronous Trigger Link mode
Semi-synchronous Trigger Link mode
Enable Source Bypass
Disable Source Bypass
IEEE-488 Reference
' Set timer for Measure Layer
' Get response from 2002
Specify Trigger Link protocol
' Select semi-sync proto-
col
' Get response from 2002
3-155