Keithley 2002 User Manual page 291

Hide thumbs Also See for 2002:
Table of Contents

Advertisement

Keithley Instruments, Inc.
28775 Aurora Road
Cleveland, Ohio 44139
1-888-KEITHLEY
www.keithley.com
Resistance Uncertainty: = ±[(ppm of reading) x (measured value) + (ppm of range) x (range used)] / 1,000,000.
% Accuracy: = (ppm accuracy) / 10,000.
1ppm of Range: = 20 counts for ranges up to 200MΩ and 10 counts on 1GΩ range at 7H digits.
SPEED AND ACCURACY
10PLC DFILT On,
RANGE
10 Readings
20 Ω
15+11+0
200 Ω
15+8+0
2 kΩ
7+0.8+0
20 kΩ
7+0.8+0
200 kΩ
29+0.8+0
2 MΩ
55+0.5+0
32
20 MΩ
175+0.6+0
32
200 MΩ
500+3+0
32
1 GΩ
2000+15+0
PLC = Power Line Cycles. DFILT = Digital Filter.
2-WIRE ACCURACY
±(ppm of range)
Additional
Uncertainty
Range
(inside T
± 5°C)
CAL
20 Ω
300 ppm
200 Ω
30 ppm
2 kΩ
3 ppm
36
NORMAL ACCURACY
RANGE
24 Hours
20 Ω
200 Ω
2kΩ
3.5+1.1
20kΩ
4.5+1.1
200kΩ
11+1.1
2MΩ
27+0.9
32
20MΩ
75+0.2
32
200MΩ
200+1
32
1GΩ
1250+5
Settling Characteristics: Pre-programmed settling delay times are for <500pF external circuit capacitance. Reading settling times are affected
by source impedance and cable dielectric absorption characteristics.
Ohms Voltage Drop Measurement: Available as a multiple display.
Autoranging: Autoranges up at 105% of range, down at 10% of range.
2-WIRE RESISTANCE READING RATES
Measurement
PLC
Aperture
Bits
10
167 ms (200 ms)
29
2
33.4 ms (40 ms)
27
1
16.7 ms (20 ms)
26
37
0.2
3.34 ms (4 ms)
23
37
0.1
1.67 ms (2 ms)
22
37
0.02
334 µs (400 µs)
20
37
0.01
167 µs (167 µs)
19
37
0.01
167 µs (167 µs)
19
36
Specifications are for 1 power line cycle, normal autozero, digital filter off, autorange off, 4-wire mode, offset compensation off, except as noted.
37
Ohms measurements at rates lower than 1 power line cycle are subject to potential noise pickup. Care must be taken to provide adequate shielding.
Specifications are subject to change without notice.
SPEC-2002 Rev. H / February 2009
90 Days
±(ppm of reading+ppm of range+ppm of range rms noise
10PLC
DFILT Off
15+11+0.5
15+8+0.5
7+0.8+0.05
7+0.8+0.1
29+0.8+0.1
53+0.5+0.1
175+0.6+0
510+3+0
2100+15+0
Temperature
Coefficient
(outside T
±5°C)
CAL
70 ppm/°C
7 ppm/°C
0.7 ppm/°C
1PLC, Offset comp. off, DFILT off
Relative Accuracy
± (ppm of reading + ppm of range)
35
7
90 Days
5+12
15+16
7+8
17+11
8+1.4
9+1.4
34+1.4
68+1.1
200+0.6
550+3
2500+15
12, 17
Default
Readings/Second to Memory
Digits
Autozero Off
Auto zero On
6 ( 5 )
29 (25)
56 (48)
222 (197)
330 (317)
330 (330)
384 (384)
2000 (2000)
11, 36
Accuracy
1PLC DFILT On,
10 Readings
DFILT Off
15+13+0.5
15+13+1
17+8+0.5
17+8+1
8+0.8+0.07
8+0.8+0.2
8+0.8+0.1
9+0.8+0.2
31+0.8+0.1
34+0.8+0.2
58+0.5+0.1
68+0.5+0.2
175+0.6+0
200+0.6+0
510+3+0
550+3+0
2100+15+0
2500+15+0
7
1 Year
17+17
19+12
10+1.5
11+1.5
40+1.5
80+1.1
275+0.6
600+3
2550+15
Readings/Second to IEEE-488
Autozero Off
2 (1.7)
6 (5)
9 (7.6)
29 (24)
47 (40)
55 (45)
156 (139)
220 (196)
176 (169)
305 (293)
182 (182)
305 (305)
186 (186)
352 (352)
2000 (2000)
Multimeter Specifications
12
)
37
1PLC
0.1PLC
DFILT Off
15+16+25
17+10+15
8+1+2
40+1+2
250+1+2
750+0.7+2
Temperature Coefficient
± (ppm of reading + ppm of range)/°C
7
Outside T
2 Years
20+19
22+13
12+1.6
13+1.6
45+1.6
90+1.1
325+0.6
650+3
2600+15
Readings/Second with
18
Time Stamp to IEEE-488
Autozero On
Autozero Off
2 (1.6)
6 (5)
9 (7.4)
27 (22)
46 (38)
50 (41)
148 (132)
156 (139)
166 (159)
157 (151)
172 (172)
160 (160)
172 (172)
179 (179)
Model 2002
37, 13
0.01PLC
DFILT Off
110+200+35
110+200+35
130+230+5
130+230+5
± 5°C
CAL
2.5+2.5
2.5+1.8
0.8+0.18
0.8+0.18
3.5+0.18
7+0.1
20+0.1
80+0.5
400+2.5
18
Autozero On
2 (1.6)
9 (7.4)
42 (34)
107 (95)
110 (106)
113 (113)
123 (123)
Page 7 of 14

Advertisement

Table of Contents
loading

Table of Contents