Source <Name - Keithley 2002 User Manual

Hide thumbs Also See for 2002:
Table of Contents

Advertisement

IEEE-488 Reference
Query
Description
Program fragment
Parameters
Query
Description
Program fragment
Parameters
Query
3-150
:READ?
This command is used to specify how much of the defined buffer (:POINts) will be used to store
pre-triggered readings. This command is essentially the same as the :PERCent command except
the amount is expressed as an exact value instead of a percent.
Sending a value that exceeds the defined size of the buffer (see :POINts) generates an error. The
command is ignored.
PRINT #1, "output 16; :trac:poin; 75"
PRINT #1, "output 16; :trac:feed:pret:amo:read 30; read?"
PRINT #1, "enter 16"
:SOURce <name>
:TRACe:FEED:PRETrigger:SOURce <name>
<name> = EXTernal
TLINk
BUS
MANual
:SOURce? Query buffer source.
This command is used to specify the event for the pre-trigger storage mode. With the pre-trigger
mode selected (:CONTrol, PRETrigger), readings are continuously stored in the buffer while
waiting for the pre-trigger source event.
With EXTernal source selected, an external trigger via the rear panel EXTERNAL TRIGGER
Input provides the pre-trigger event.
With TLINk selected, an input trigger via the Trigger Link provides the pre-trigger event. The
Trigger Link line used for the input trigger is programmed from the Trigger subsystem (see
:TRIGger:TCONfigure command path).
With BUS selected, a GPIB bus trigger (GET or *TRG) provides the pre-trigger event.
With MANual selected, pressing the front panel TRIG key provides the pre-trigger event.
PRINT #1, "output 16; :trac:feed:pret:sour bus; sour?"
PRINT #1, "enter 16"
:CONTrol <name>
:TRACe:FEED:CONTrol
<name>
<name> = NEVer
NEXT
ALWays
PRETrigger
:CONTrol? Query buffer control
Specify pre-trigger event
Use External Trigger as pre-trigger event
Use Trigger Link as pre-trigger event
Use bus trigger as pre-trigger event
Use TRIG key as pre-trigger event
Specify buffer control
Disables buffer storage
Fills buffer and stops
Continuously stores readings in buffer
Use Pretrigger mode to store readings
' Set buffer size
' 30
pre-trugger
rdgs for buffer
' Get response from
2002
' Select bus pre-trig-
ger
' Get
response
from
2002

Advertisement

Table of Contents
loading