Keithley 2002 User Manual page 175

Hide thumbs Also See for 2002:
Table of Contents

Advertisement

IEEE-488 Reference
Table 3-10 (Continued)
SENSe command summary
Command
:CURRent:DC
:APERture <n>
:AUTO <b>
:AUTO ONCE
:AUTO?
:APERture?
:NPLCycles <n>
:AUTO <b>
:AUTO ONCE
:AUTO?
:NPLCycles?
:RANGe
[:UPPer] <n>
[:UPPer]?
:AUTO <b>
:AUTO ONCE
:ULIMit <n>
:ULIMit?
:LLIMit <n>
:LLIMit?
:AUTO?
:REFerence <n>
:STATe <b>
:STATe?
:ACQuire
:REFerence?
:DIGits <n>
:AUTO <b>
:AUTO ONCE
:AUTO?
:DIGits?
:AVERage
:TCONtrol <name>
:TCONtrol?
:COUNt <n>
:COUNt?
:ADVanced
:NTOLerance <n>
:NTOLerance?
[:STATe] <b>
[:STATe]?
[:STATe] <b>
[:STATe]?
:AUTO <b>
:AUTO ONCE
:AUTO?
:METHod <name>
:METHod?
3-52
Description
Path to configure DC current.
Set integration rate in seconds (166.67e-6 to 1).
Enable or disable auto aperture.
Enable and then disable auto aperture.
Query auto aperture.
Query aperture (integration rate).
Set integration rate (line cycles; 0.01 to 50).
Enable or disable auto NPLC.
Enable and then disable auto NPLC.
Query auto line cycle integration.
Query line cycle integration rate.
Path to configure measurement range:
Select range (0 to 2.1).
Query range.
Enable or disable auto range.
Set range based on present input signal.
Set upper limit for auto range (0 to 2.1).
Query upper limit.
Set lower limit for auto range (0 to 2.1).
Query lower limit.
Query auto range.
Specify reference (-2.1 to +2.1).
Enable or disable reference.
Query state of reference.
Use input signal as reference.
Query reference value.
Specify measurement resolution (4 to 9).
Enable or disable auto resolution.
Enable and then disable auto resolution.
Query auto resolution (0 or 1).
Query resolution.
Path to configure and control the filter.
Select filter type: (MOVing or REPeat).
Query filter type.
Specify filter count (1 to 100).
Query filter count.
Path to configure and control advanced filter.
Specify noise tolerance level (0 to 100%).
Query noise tolerance level.
Enable or disable advanced filter.
Query state of advanced filter.
Enable or disable filter.
Query state of digital filter.
Enable or disable auto filter.
Enable and then disable auto filter.
Query auto filter.
Select measure technique: (NORMal or ICIRcuit).
Query method.
Default
parameter
SCPI
Ref.
(Note 2)
3.18.5
OFF
1
3.18.6
OFF
3.18.8
2.1
ON
2.1
0.0002
0
3.18.9
OFF
7
3.18.10
ON
3.18.11
(Note 3)
10
1
ON
(Note 4)
(Note 5)
3.18.14
NORMal

Advertisement

Table of Contents
loading

Table of Contents