Pwindow <N; Method <Name; Source <Name - Keithley 2002 User Manual

Hide thumbs Also See for 2002:
Table of Contents

Advertisement

IEEE-488 Reference
3.18.13
Parameters
Query
Description
Program fragment
3.18.14
Parameters
Query
Description
Program fragment
3.18.15
Parameters
Query
3-106
:PWINdow <n>
[:SENSe[1]]:VOLTage:AC:DETector:PWINdow <n>
<n> = 0.1 to 9.9
Peak spike detection window (in seconds)
DEFault
0.1sec window
MINimum
0.1sec window
MAXimum
9.9sec window
:PWINdow?
:PWINdow? DEFault
:PWINdow? MINimum
:PWINdow? MAXimum
This command is used to set the detection window for peak spike measurements (:NPEak or
:PPEak). The instrument measures and displays the largest peak spike that occurs within each
specified detection window. For example, for a
peak spike after every
second. See paragraph 2.4.1 (AC-TYPE; Positive Peak and Negative
Peak) for more information.
PRINT #1, "output 16; :volt:ac:det:pwin 1; pwin?"
PRINT #1, "enter 16"
:METHod <name>
[:SENSe[1]]:CURRent:DC:METHod <name>
<name> = NORMal
Normal DCI measurement technique
ICIRcuit
In-circuit DCI measurement technique
:METHod? Query measurement technique
This command is used to select the DCI measurement technique. NORMal is used to make con-
ventional current measurements which are performed by breaking the circuit and placing the
ammeter in series with it.
The ICIRcuit selection allows you make an in-circuit current measurement of a wire or circuit
board trace. This measurement technique is performed by making four-wire INPUT connections
to the wire or trace (the AMPS input is not used). The instrument automatically performs a four-
wire ohms offset compensated measurement and calculates the current. See paragraph 2.4.2
(IN-CIRCUIT) for details.
PRINT #1, "output 16; :curr:dc:meth icir; meth?"
PRINT #1, "enter 16"
:SOURce <name>
[:SENSe[1]]:FREQuency:SOURce <name>
<name> = CURRent
Select current frequency input source
VOLTage
Select voltage frequency input source
:SOURce? Query frequency input source
Set peak detection window for ACV
Query spike detection window
Query *RST default window
Query lowest allowable window
Query largest allowable window
second window, the instrument displays the
Specify measurement technique for DCI.
Specify input source for FREQ.
' Specify 1 sec window
' Get response from 2002
' Select in-circuit I
measure
' Get response from 2002