Keithley 2002 User Manual page 256

Hide thumbs Also See for 2002:
Table of Contents

Advertisement

Program fragment
Effects of negative transitions on the Questionable Event Register:
Questionable event
Temperature summary
Calibration summary
Command warning
Effects of negative transitions on the Operation Event Register:
Negative transition effect on
Operation event
Operation Event Register
Calibrating
Sets B0 at the end of calibration.
Measuring
Sets B1 at the end of a measurement.
Trigger layer
Sets B5 when leaving the Trigger Layer.
Arm layer
Sets B6 when leaving an arm layer.
Idle
Sets B10 when leaving the idle state.
Effects of negative transitions on the Trigger Event Register:
Negative transition effect on
Trigger event
Trigger Event Register
Sequence 1
Sets B1 when leaving the Trigger Layer.
Effects of negative transitions on the Arm Event Register:
Negative transition effect on
Arm event
Arm Event Register
Sequence 1
Sets B1 when leaving an arm layer.
Effects of negative transitions on the Sequence Event Register:
Negative transition effect on
Sequence event
Sequence Event Register
Layer 1
Sets B1 when leaving arm layer 1.
Layer 2
Sets B2 when leaving arm layer 2.
PRINT #1, "output 16; :stat:oper:ntr 1026; ntr?"
PRINT #1, "enter 16"
Negative transition effect on
Questionable Event Register
Sets B4 when a valid reference junction measurement follows an
invalid reference junction measurement.
Sets B8 when an invalid calibration constant clears after a success-
ful calibration.
Sets B14 when a Signal Oriented Measurement Command parame-
ter executes correctly after a previous Signal Oriented Measurement
Command parameter has been ignored.
IEEE-488 Reference
' Set B1 and B10 of Operation
NTR
' Get response message from
2002
3-133

Advertisement

Table of Contents
loading