Parameters
Query
These commands are used to enable or disable LIMIT 1 and LIMIT 2 tests. When enabled, the
test sequence for limits will be performed every time the instrument performs a measurement.
Testing is performed in the following sequence: Low Limit 1, High Limit 1, Low Limit 2 and
High Limit 2. Any limit test (LIMIT 1 or LIMIT 2) not enabled is simply not performed.
Note that when a limit test (LIMIT 1 or LIMIT 2) is enabled, the digital output port cannot be
controlled from the Source Subsystem.
A fail indication (see :FAIL?) for LIMIT 1 or LIMIT 2 is cleared when the respective limit test
is disabled.
PRINT #1, "output 16; :calc3:lim:stat on; stat?"
PRINT #1, "enter 16"
2002
:FAIL?
:CALCulate3:LIMit[1]:FAIL?
:CALCulate3:LIMit2:FAIL?
These commands are used to read the results of LIMIT 1 and LIMIT 2 tests:
0 = Limit test passed
1 = Limit test failed
The response message (0 or 1) only tells you if a limit test has passed or failed. It does not tell
you which limit (upper or lower) has failed. To determine which limit has failed, you will have
to read the Measurement Event Register (see paragraph 3.20.1).
Reading the results of a limit test does not clear the fail indication of the test. A failure can be
cleared by using a :CLEar command, or by disabling the test (:STATe OFF).
PRINT #1, "output 16; :calc3:lim:fail?"
PRINT #1, "enter 16"
:CLEar commands
[:IMMediate]
:CALCulate3:LIMit[1]:CLEar[:IMMediate]
:CALCulate3:LIMit2:CLEar[:IMMediate]
These action commands are used to clear the fail indication of LIMIT 1 and LIMIT 2 tests. Note
that a failure is also cleared when the limit test is disabled (:STATe OFF).
PRINT #1, "output 16; :calc3:lim:cle"
:AUTO <b>
:CALCulate3:LIMit[1]:CLEar:AUTO <b>
:CALCulate3:LIMit2:CLEar:AUTO <b>
<b> = 1 or ON
Enable auto-clear for limit failure
0 or OFF
Disable auto-clear for limit failure
:AUTO? Query state of auto clear
' Enable LIMIT 1 test and
query
' Get response message from
Read LIMIT 1 test results
Read LIMIT 2 test results
' Query result of LIMIT 1 test
' Get response message from 2002
Clear LIMIT 1 test failure
Clear LIMIT 2 test failure
! Clear fail condition of LIMIT 1 test.
Control auto-clear; LIMIT 1
Control auto-clear; LIMIT 2
IEEE-488 Reference
3-73