Description
Example
OCOMP
Purpose
Format
Parameter
Default
Query
Description
Example
OFORMAT
Purpose
Format
Parameter
Default
Query
The NRDGS command programs the number of readings per trigger as well as the event that initiates
the trigger. Available events include continuous, external trigger, the GPIB talk command, as well as
timer events.
All other HP3458A trigger events are not supported by the Model 2002.
NRDGS 10, SYN
NRDGS 100, EXT
To enable or disable the offset-compensated ohms mode.
OCOMP <control>
<control>
<control> parameter
Numeric equivalent
OFF
0
ON
1
ON (power-on default = OFF)
Returns the current offset-compensated ohms state.
OCOMP?
The OCOMP command enables/disables the offset-compensated ohms function.
Enable offset-compensated ohms.
OCOMP ON
To program the data format for both normal and memory readings.
OFORMAT <format>
<format>
<format> parameter
Numeric equivalent
ASCII
1
SREAL
4
DREAL
5
ASCII
Returns the present setting of the output format.
OFORMAT?
Take 10 readings per talk command.
Take 100 readings per external trigger event.
Description
Offset-compensated ohms off
Offset-compensated ohms on
Description
ASCII
Single real
Double real
HP3458A Emulation Mode