Function] <Name - Keithley 2002 User Manual

Hide thumbs Also See for 2002:
Table of Contents

Advertisement

Program fragment
Parameters
Query
Description
Program fragment
3.18.12
Parameters
Query
Description
Program fragment
PRINT #1, "output 16; :volt:dc:aver:adv:ntol 20; ntol?"
PRINT #1, "enter 16"
[:STATe] <b>
[:SENSe[1]]:CURRent:AC:AVERage:ADVanced[:STATe] <b>
[:SENSe[1]]:CURRent:DC:AVERage:ADVanced[:STATe] <b>
[:SENSe[1]]:VOLTage:AC:AVERage:ADVanced[:STATe] <b>
[:SENSe[1]]:VOLTage:DC:AVERage:ADVanced[:STATe] <b>
[:SENSe[1]]:RESistance:AVERage:ADVanced[:STATe] <b>
[:SENSe[1]]:FRESistance:AVERage:ADVanced[:STATe] <b>
<b> = 0 or OFF
Disable advanced filter
1 or ON
Enable advanced filter
[:STATe]? Query state of advanced filter
These commands are used to enable or disable the advanced filter for the specified function. The
:ADVanced:NTOLerance commands are used to set tolerance levels.
PRINT #1, "output 16; :volt:dc:aver:adv on; adv?"
PRINT #1, "enter 16"
:DETector commands
[:FUNCTION] <name>
[:SENSe[1]]:CURRent:AC:DETector[:FUNCTION] <name>
[:SENSe[1]]:VOLTage:AC:DETector[:FUNCTION] <name>
ACI:
<name> = RMS
AVERage
ACV:
<name> = RMS
AVERage
LFRMs
PEAK
NPEak
PPEak
[:FUNCtion]? Query selected measurement type
These commands are used to select the measurement type for the ACI and ACV functions. Refer
to paragraphs 2.4.1 (ACV) and 2.4.2 (ACI) for details on AC-TYPE.
PRINT #1, "output 16; :curr:ac:det aver; det?"
PRINT #1, "enter 16"
AC RMS amps measurements
AC average amps measurements
AC RMS volts measurements
AC average volts measurements
Low frequency AC volts measurements
AC peak volts measurements
AC negative peak spike volts measurements
AC positive peak spike volts measurements
IEEE-488 Reference
' Get noise
tolerance
' Get response from
2002
Control advanced filter for ACI
Control advanced filter for DCI
Control advanced filter for ACV
Control advanced filter for DCV
Control advanced filter for
Control advanced filter for
' Enable advanced filter
' Get response from 2002
Specify detector for ACI
Specify detector for ACV
' Select average detector
' Get response from 2002