HP3458A Emulation Mode
Query
Description
Example
DISP
Purpose
Format
Parameter
Default
Query
Description
Example
EMASK
Purpose
Format
H-6
Returns delay time in seconds.
DELAY?
The DELAY command programs the time interval between the trigger event and the first sample event
in seconds. Although the valid parameter range is from 1E-7s to 6000s, the Model 2002 rounds any
value below 1ms (0.001s) to 1ms. Specifying a delay of 0 selects the minimum delay time possible.
Program 50ms delay.
DELAY 50E-3
To enable/disable the front panel display, to clear the display, and to send a message to the display.
DISP <control>, "<message>"
<control>
<control> Parameter
OFF
ON
MSG
CLR
Up to 52-character string enclosed in double quotes.
<message>
ON
Returns present value of <control> parameter.
DISP?
The DISP command enables/disables the multimeter's display and may be used to send a message to
the display, or to clear the display. ON selects normal operation. OFF displays a message if a valid
message parameter has been sent; otherwise, all dashes are displayed. Readings are no longer dis-
played, and the display is not updated, except to service front panel keystrokes. The MSG parameter
simply displays a message on the front panel, and the CLR parameter clears the display entirely. The
parameter is any string up to 52 characters enclosed in double quotation marks.
<message>
DISP OFF
DISP ON
DISP MSG, "Model 2002"
To enable certain error conditions to set the error bit in the status register.
EMASK <bit_value>
Numeric equivalent
0
1
2
3
Turn off display.
Enable normal display operation.
Display message.
Description
Display off or message.
Normal operation.
Display message.
Clear display.